ASTM D525 Automatic Oxidation Stability Of Gasoline Induction Period Method Tester equipment analyzer
Negotiable /Set
Min.Order:1 Set
Hunan Addition Instruments and Apparatus Co., Ltd.
Automatic Oxidation Stability Of Gasoline
(Induction Period Method) Tester ASTM D525
Product model: JF8018Z
Product Introduction
This instrument measures the oxidation stability of gasoline under accelerated oxidation conditions in accordance with GB/T8018, ISO5736, ASTM D525, DIN51781 "Determination of Gasoline Oxidation Stability (Induction Period Method)". The induction period can be used to indicate the tendency of gasoline to form gums during storage.
Working conditions and environment
There is no corrosive gas in the room; the ambient temperature is 5-35℃; the power supply has a good grounding.
The main technical parameters
1 Working voltage: 198-242V
2 Power frequency: 49-51Hz
3 Heater power: 2000W
4 Constant temperature mode: automatic control
5 Test temperature range: 98~102℃
6 Temperature control: ±0.1℃
7 Temperature measuring element: platinum resistance (Pt100)
8 Thermometer graduation: 0.1℃
9 Bathtub capacity: 28.8 l Bathtub depth: 320mm
10 Pressure gauge: 0~1.6MPa, 0.4 grade
11 Mixing method: electric
Structural features:
1 This tester is a split desktop machine with compact structure, convenient operation and comfortable.
2 It can hold two GB/T8018 standard oxygen bombs at the same time. The oxygen bomb body has good sealing performance and is easy to assemble and disassemble.
3 The bathtub material is stainless steel, which is rust-proof and durable.
4 This tester directly adopts the upper cover of the oxidation bath for condensation. Because the oxidation bath is airtight and has good condensing performance, there is no need for refilling in one test.
5 The temperature is adjusted by time proportional type, and the heating power can be adjusted in the range of 0~2000W.