PPG-2R-880 / PPG-2R-940 Reflectance SpO2 Module

Brand Name whaleteq
Place of Origin Taiwan
Model Number PPG-2R-880 / PPG-2R-940 Reflectance SpO2 Module
Power Other, 100 ~ 3000mV
Usage Other, reflectance SpO2 performance tester

Product Features


PPG-2R-880 / PPG-2R-940 Reflectance SpO2 Module

The PPG-2R-880/PPG-2R-940 works with the AECG100 main console and software, as the first reflectance SpO2 performance tester designed for R&D in manufacturers. This dual-channel optical module outputs R and IR analog signals to test blood oxygen saturation (SpO2) and heart rate functions. Besides, it also allows users to individually adjust the R/IR’s rate of change of respiratory signals in PPG and set up add-on noise. It’s suitable for the development and verification of reflectance PPG products such as wearable devices to test performance.

  • Works with the AECG100 main console and software to test heart rate, ECG, PPG, PWTT, and SpO2 functions

  • Outputs R and IR analog signals to simulate SpOsignals

  • Adjustable SpO2 parameters, such as AC, DC, and PI values of R/IR, to confirm the accuracy of the SpO2 function

  • Adjustable PPG waveform parameters (R/IR amplitude, time, and heart rate) in a wide range and precise values for flexible test settings

  • ECG function with adjustable PWTT parameters to verify the accuracy of blood pressure measurement algorithms

  • Built-in 7 types of noise and respiratory variation simulation - add-on signals for more authentic testing

  • Equipped with Auto Sequence function - sorts the parameters of different test items into a designated order for automated testing while eliminating the need to develop automated testing programs

  • Raw data playback - plays recorded raw data or customized waveforms for repeated verification to facilitate the development process

  • Provides SDK supporting multiplatform for users to flexibly develop automated testing programs

  • Equipped with an RCA to BNC port to connect to an oscilloscope - outputs amplified and stable signals for viewing the voltage of AC analog signals of R/IR, optical signals received by the PPG-2R-880/PPG-2R-940, and the switch status of R/IR LED of DUT

  • ParametersSpecifications
    Heart RateRange10 ~ 300BPM
    Resolution1BPM
    Accuracy±1BPM
    LED DC Level Reference OutputRange100 ~ 3000mV
    Resolution1mV



    LED AC Level Reference OutputRange0.75 ~ 30mV
    Resolution0.01mV



    PI (AC level / DC level)Range0.025% ~ 30% (AC / DC, varies with AC or DC level)
    AccuracyNA



    LED Light WavelengthRangeRed : 660nm ; Infrared : 940nm / 880nm
    AccuracyRed: ±10nm ; Infrared : ±10nm



    Red Light Absolute Irradiance (100%)§Intensity3.55mW / m^2
    Accuracy±15%



    Infrared Light Absolute Irradiance (100%)§Intensity6.65mW / m^2
    Accuracy±15%



    PD Response TimeRising1μs Typical*
    Falling1μs Typical*



    SpO2 %Range1% ~ 100%
    Resolution1%
    Accuracy±1% + specified accuracy of the DUT (using Masimo Radical-7 Pulse CO-Oximeter or "Covidien" Nellcor Portable SpOPatient Monitoring System)

    §Data Collection Method: To connect the spectrometer with a 2-inch integrating sphere, and fix the integrating sphere on the DUT to measure the absolute irradiance of the LED of the DUT.

    *Note: PD response time varies with the light intensity of the DUT.