Product Details

WECG400 ECG performance tester USING production line

Brand Name whaleteq
Place of Origin Taiwan
Model Number WECG400
Power Other, ±1% for amplitudes of 0.5mVpp or higher
Usage Other, ECG performance testing

Product Features

WECG400, Parallel Testing ECG Tester

The first parallel testing ECG tester designed for production lines can simultaneously test ECG function on up to 4 wearable devices with a single computer to accelerate testing time.
This tester provides IEC 60601-2-47 test items and outputs 7 types of standard ECG waveforms to assist DUTs in conforming with regulatory requirements.
For customized testing, the raw data playback function can reproduce clinical data or user-defined waveforms to help users verify that the DUTs meet the manufacturer's specifications. This tester integrates needed equipment to optimize testing efficiency while ensuring product quality and a stable production process. It’s suitable for manufacturers who have mass-production needs on wearable devices with the ECG function.

Compliance tests conforming with IEC 60601-2-47, China YY0885, and YY 9706.247-2021

  • Integrates IEC 60601-2-47 test items*, including dynamic range, input impedance, frequency response, etc.

  • Equipped with 4 sets of electrode terminals - RA/LA/RLD electrode terminals on each set to meet testing needs for wearable device's ECG function

  • Outputs 7 types of standard ECG waveforms - ECG Waveform, Sine, Triangle, Square, Exponential, Rectangle Pulse, and Triangle Pulse

  • Provides SDK - users can flexibly develop automated testing programs and control multiple WECG400 to test more than 4 DUTs simultaneously

  • Supports sequential testing - tests DUTs one at a time according to a user-defined sequence

  • Equipped with 4 DUT Monitor ports to connect to an oscilloscope - amplifies the ECG signals to help users monitor the signal output with higher resolutions

  • Rack-mountable and easy to deploy with the 1U height design.

  • ParametersRangeResolutions
    ECG WaveformFrequency (Hz)0.02 ~ 6Hz0.01
    Frequency (BPM)1 ~ 360BPM1
    Amplitude(-5) ~ 5mV0.01
    SineFrequency (Hz)0.05 ~ 500Hz0.01
    Frequency (BPM)3 ~ 30000BPM1
    Amplitude(-10) ~ 10mV0.01
    TriangleFrequency (Hz)0.05 ~ 500Hz0.01
    Frequency (BPM)3 ~ 30000BPM1
    Amplitude(-10) ~ 10mV0.01
    SquareFrequency (Hz)0.05 ~ 500Hz0.01
    Frequency (BPM)3 ~ 30000BPM1
    Amplitude(-10) ~ 10mV0.01
    ExponentialFrequency (Hz)0.05 ~ 5Hz0.01
    Frequency (BPM)3 ~ 300BPM1
    Amplitude(-5) ~ 5mV0.01
    Rectangle PulseFrequency (Hz)0.05 ~ 5Hz0.01
    Frequency (BPM)3 ~ 300BPM1
    Amplitude(-5) ~ 5mV0.01
    Pulse width (frequency: 0.05Hz)1 ~ 300ms1
    Pulse width (frequency: 5Hz)1 ~ 200ms1
    Triangle PulseFrequency (Hz)0.05 ~ 5Hz0.01
    Frequency (BPM)3 ~ 300BPM1
    Amplitude(-5) ~ 5mV0.01
    Pulse width (frequency: 0.05Hz)1 ~ 300ms1
    Pulse width (frequency: 5Hz)1 ~ 200ms1

     

  • Input Impedance / Frequency Response Signal add-on
  •              
    ParametersRangeDefault ValuesResolutions
    Play raw dataSample rateMaximum: 40K Hz--
    DC offset (non-Variable)-300, 0, +300mV0mV300
    DC offset (Variable)(-1000) ~ 1000mV0mV100
    Input impedance switch100Ω (short circuit)--
    620kΩ / 4.7 nF on/offOff (on: short circuit)-
    10MΩ / 180 pF on/offOff (on: short circuit)-

     

  • Product System Capacity and Standard Requirements Table
  • ParametersSpecificationsSystem Capability
    Main output voltage accuracy±1% for amplitudes of 0.5mVpp or higher±0.3%
    Main output voltage resolution
    (DAC resolution)
    N/A0.15µV
    Frequency / pulse repetition rate accuracy±1%±0.1%
    Pulse duration / timing accuracy±1ms±0.2ms
    Resistor tolerance±1%±0.5%
    Capacitor tolerance±5%±5%
    Precision 1000 : 1 divider±0.5%±0.3%
    Sample rate5kHz ± 0.1%10kHz ± 0.05% (50ppm)
    DC offset
    (fixed, noise-free, sourced from the internal supercapacitor)
    300mV ± 1%300mV ± 0.1%
    DC offset
    (variable, up to 1000mV)
    Setting ±1% or ±3mVSetting ±1% or ±3mV
    Environment15 ~ 30°C (by design)
    30 ~ 80% RH (by design)
    5 ~ 40°C
    30 ~ 80% RH
    altitude <2000M
    Overvoltage CAT II