FDA CE Nitrile Examination Glove
USD $38 - $42 /Carton
Min.Order:50 Cartons
Nitrile Glove Powder Free (Non-sterile)
-Premium Quality (1st Grade)
-Ideal for healthcare, medical used, food processing and general industrial used.
- FDA (Food safe certified).
# | Characteristics | Inspection Level | Acceptable Quality Level | Reference Standard |
2.1 | Dimensions | S2 | 4.0 | ASTM D6319-00ae3 |
2.2 | Physical Properties | S2 | 4.0 | ASTM D6319-00ae3 |
2.3 | Freedom from holes (Air Pump Test) | GI | 1.5 | In-house practice |
2.4 (i) (ii) | Visual Defects: - Major Visual - Minor Visual | GI |
1.5 2.5 | In-house practice |
2.5 (i) (ii) (iii) | Packaging Defects: Regulatory Visual Critical (incl. Gloves Counting) |
GI GI S2 |
** 4.0 4.0 |
In-house Inspection |
2.6 | Powder Free Residue | N=5 | - | ASTM D6319-00ae3 ASTM D6124-06 |
2.7 | Mix Size / Mix Glove / Mix Hand | Not Allowed |
Description | Size | Standard |
Length (mm) | All Sizes | Min 240 |
Palm Width (mm) | XS | 76 +/- 3 |
| S | 84 +/- 3 |
| M | 94 +/- 3 |
| L | 105 +/- 3 |
| XL | 113 +/- 3 |
Thickness (mm) Single wall | All Sizes |
Finger : 0.12 +/- 0.02
(Typical value: 0.13 – 0.14)
Palm : 0.10 +/- 0.02
(Typical value: 0.10 – 0.12)
|
Description | Standard | |
Before Aging | After Aging | |
Elongation at break (%) | min 500 (Typical value: 500 - 600) | min 400 (Typical value: 400 - 500) |
Tensile Strength (MPa) | min 14 (Typical value: 14 - 18) | min 14 (Typical value: 14 - 18) |
3.1 Freedom fromholes
The sample size and allowablenumber of non-conforming gloves in the samples shall be determined in accordance to Sampling Plan ISO 2859-1 Single Normal using inspection and acceptable quality level as stated in Section II: Performance Requirements.
The sample size and allowable number of non-conforming gloves in the samples for both major and minor defects shall be determined in accordance to Sampling Plan ISO 2859-1 Single Normal using Inspection and acceptable quality level as stated in Section II: Performance Requirements.
The Sample size and allowable number of non-conforming in sample for regulatory, visualand critical packaging defects shall be determined in accordance to Sampling Plan ISO 2859- 1 Single Normal using inspection and acceptable quality level as statedin Section II: Performance Requirements (Gloves Counting = 100 pcs by weight per Dispenser).
Maximum 2 mg per glove