- GOAL: check that all the touch accessible parts are properly connected to the protective earth connection.
- PRINCIPLE: Apply a current in between each metallic accessible part and the earth protection connection, then measure the voltage in between those two parts and check with the Ohmic law that the calculated resistance is lower than the value required by the standard. The DUT is not connected to mains!
- TEST CONDITIONS (according to the main standards):
♦ Current 10, 25 A AC or DC, or 2 x In
♦ Voltage from 6 to 12 volts AC or DC
♦ Resistance < 0, 1ohm, 0, 2 ohm or 0, 5 ohm
♦ Time > 5 s, 1mn or not defined
Combined with CS9911N series hipot tester, CS9950N can do the hipot test simultaneously, which saves the test time. Our company provide the automatic test box.
The above picture shows the ACW hipot, DCW hipot, insulation resistance and ground resistance test simultaneously.
- TECHNICAL PARAMETERS
Model | CS9950N | CS9950AN | ||
GR (ground resistance)
| Output current | range | (5.0~32.0)A | (5.0-25.0)A |
accuracy | ±(2%read+2counts) | |||
resolution | 0.1A | |||
Maximum output power | 153.6VA | 93.75VA | ||
Upper limit set |
|
| ||
Output waveform | Sine waveform | |||
Waveform distortion | ≤2%( Unloaded or pure resistance of the load) | |||
Crest factor | 1.3~1.5 | |||
Signal type | DDS+ Amplifier | |||
Test time | 0.3s~999.9s 0=continuous | |||
Interval time | 0.0s~999.9s 0=OFF | |||
Resistance offset | 0~10mΩ | |||
AMPERE METER (G.R.) | Range | (5.0~32.0)A | (5.0-25.0)A | |
Resolution | 0.1A | |||
Accuracy | ±(2%+2counts) | |||
Display | RMS | |||
OHM METER (G.R.) | Range | 0~510mΩ | ||
Accuracy | ±(2%+2mΩ) | |||
Resolution | 1mΩ | |||
Measurement | Four terminal method | |||
TIMER | Range | 0~999.9s | ||
Resolution | 0.1s | |||
Accuracy | ±(1%+50ms) | |||
Memory group | 20 | |||
Test step | 8 |