LMQ.R-3870 Vertical Autoclave
Negotiable /Set
Min.Order:1 Set
SHINVA Medical Instrument Co., Ltd. (Shandong)
Major capability and characteristics :
PLC control technique, touch-key.
Automatic control over the processes of heating, sterilization, steam exhaust and dryness.
Parameters of program can be designed by user accord to their requirements.
Process, temperature, time dynamic LCD displayed. Fault automatic displayed and alarm.
Positive pressure pulse function exhaust cool air of sterilizing chamber thoroughly and ensure the steam saturation.
The buzzer alarms when a cycle of operation is completed.
Preheating control function.
Built-in water tank. Steam and water inner cycle system, no external water source, not exhaust steam outside, environment is clean and dryness.
Horizontal sliding door, self- bulge seal ring, operated easily.
Safety devices:
Auto protection device against over temperatureThis system cuts off power automatically if the designed temperature is surpassed. Light and sound alarm operator meanwhile.
The program will be stopped if water level lower than the low limit, alarm with sound and light.
Door safety interlockThe door cannot be opened until there is no pressure in the chamber and power for heating cannot be switched on until the door is firmly closed.
Safety valve of auto release against overpressureIt opens when the designed pressure is surpassed.
Heat insulation dock shelter: the door is made of high performance heat insulation material, not only beautiful but also prevents the operator to be injured.
Over current and creepage protector.
Main technical parameters:
Designed pressure: 0.26Mpa
Designed temperature: 150
Rated working pressure: 0.22Mpa
Sterilization temperature: 105~136
Sterilization time range: 0~99min
Material of chamber: SUS 304 stainless steel
Accessories: two stainless steel sterilization baskets