CE/ISO certified; TMQ.J-2540 Autoclave
Negotiable /Set
Min.Order:1 Set
SHINVA Medical Instrument Co., Ltd. (Shandong)
Major capability and characteristics :
Manual fill water, drying by exhaust steam.
Automatic control over the processes of heating and sterilization.
Imported pressure switch, automatic pressure control.
Steam and water circulation system, no external water source, not exhaust steam outside, environment is clean and dryness.
Safety devices:
Auto protection device against over temperatureThis system cuts off power automatically if the designed temperature is surpassed.
Door safety interlockThe door cannot be opened until there is no pressure in the chamber and power for heating cannot be switched on until the door is firmly closed.
Safety valve of auto release against overpressureIt opens when the designed pressure is surpassed.
Main technical parameters:
Designed pressure: 0.26Mpa
Designed temperature: 150
Rated working pressure: 0.22Mpa
Sterilization temperature: 105~134
Pressure control precision: 0.005Mpa
Temperature degree of uniformity: ≤±0.5
Material of chamber: SUS 304 stainless steel