CE/ISO certified; TMQ.R-2540 Autoclave
Negotiable /Unit
Min.Order:1 Unit
SHINVA Medical Instrument Co., Ltd. (Shandong)
Major capability and characteristics :
PLC control technique, touch-key.
Automatic control over the processes of heating, sterilization, steam exhaust and dryness.
Process, temperature, time dynamic LCD displayed, fault will dynamic display and alarm.
Positive pressure pulse function exhaust cool air of sterilizing chamber thoroughly and ensure the steam saturation.
The buzzer alarms when a cycle is completed.
Steam and water circulation system, no external water source, not exhaust steam outside, environment is clean and dryness.
Pre-heating control function.
Safety devices:
Auto protection device against over temperatureThis system cuts off power automatically if the designed temperature is surpassed.
Door safety interlockThe door cannot be opened until there is no pressure in the chamber and power for heating cannot be switched on until the door is firmly closed.
Safety valve of auto release against overpressureIt opens when the designed pressure is surpassed.
The program will be stopped if water level lower than the low limit, alarm with sound and light.
Electronic circuits safety device: direct current control loop and has protection against overload and voltage surpassed, alternating current main loop has short-circuit protection device.
Main technical parameters:
Designed pressure: 0.26Mpa
Designed temperature: 150
Rated working pressure: 0.22Mpa
Sterilization temperature: 105~136
Precision of temperature displayed: 0.1
Temperature degree of uniformity: ≤±0.5
Material of chamber: SUS 304 stainless steel