Lab Portable Raman Spectrometer price
USD $8,000 - $18,900 /Piece
Min.Order:1 Piece
Shanghai Drawell Scientific Instrument Co., Ltd.
DTR2000 Portable Raman Spectrometer
1 Specifications
DTR2000 System | |
Interface | USB 2.0/LAN |
OS | Android 4.4.2 |
Battery endurance | >5 h |
Integration time | 4ms - 120s |
Supply voltage | DC 19V(+/-5%) |
Operating temperature | 5~40 ℃ |
Operating humidity | < 95% |
Size (L*W*H) | 30×22.5×13.2 cm3 |
Weight | 7 Kg |
Reliability | |
Spectral stability | σ/μ < 0.5% (COT 8 hours) |
Temperature stability | Spectral shift ≤ 1 cm-1 (10-40 ℃) |
Variation of Spectral intensity in 5 ~ 40 ℃ | <±5% |
Optics Parameter | |
Spectral region | 250-2400 cm-1 |
Spectral resolution | 6 cm-1 |
Signal-to-Noise | >200:1 (918 cm-1 of Acetonitrile,10s accumulation, 200mW) |
Incidence slit | 50 μm |
Optical system | f/4 Crossed Symmetrical Czerny-Turner |
Focal length | 98 mm for incidence and output |
Detector | |
Type | Linear Array Detector |
Detectable range | 200-1100 nm |
Effective pixel | 3648 |
Dynamic range | 1300:1 |
Pixel dimension | 8μm×200μm |
Full well capacity | 100 Ke- |
Sensitivity | 130 Photon @ 400 nm; 60 Photon @ 600 nm |
Exciting Laser | |
Center Wavelength | 785nm (+/-1nm) |
FWHM | 0.1 nm |
Output power | ≥450 mW |
Power stability | σ/μ <±0.2% |
Raman Probe | |
Working distance | 6 mm |
Rayleigh scattering resistance | OD>8 |
Numerical Aperture | 0.3 |
Aperture | 7mm |
Product display