Circulating Liquid Ring Vacuum Pumps
USD $1 - $800 /Set
Min.Order:1 Set
Zhengzhou Keda Machinery And Instrument Equipment Co., Ltd.
SHZ-95B is a once forming anti-corrosive shell with five taps new pump, which is improved on the basis of all stainless steel five taps in SHZ- C type vacuum pump. The taps can also wok alone, or paralleled to get more vacuum. It can meet the needs of larger rotary evaporators and glass reactors, which also serve many devices at the same time.
Application:
It provides vacuum condition for evaporation, distillation, crystallization, drying, distillation, filtration pressure, vacuum degassing experiments. Especially suitable for colleges and universities, research institutes, chemical, pharmaceutical, biochemical, food, pesticides, agricultural engineering, biological engineering etc industries’ scientific research, pilot and production.
Features:
(1)One machine with multi-pipe operation, where three sets (five sets )of air exhauster are fixed and can be used independently or parallelly, compared with desktop pump, swept volume is bigger, applied to large swept volume vacuum needs.
(2)Corrosion resistance, free-pollution, stable performance, low noise and durable.
(3)The tank is made from Polyvinyl chloride plastics, free-pollution; The shell adopts stainless steel material or static spring plastic.
(4)Host stainless steel movement and anti-corrosive movement models.
(5)The water-line target, light shows high, medium and low.
(6)4 trundles make the movement easily
(7)The air exhaustor, three links, check valve, suction nozzle are made of PTFE, anti-acid and alkali.
(8)Pump body and impeller are made of corrosion protection polypropylene.
(9)Special customized motor by well-known manufacturers, fluorine rubber mechanical seal to keep no corrosive gas into the motor.
(10)There are several sets pumping system inside the pump. Three or five sets taps are installed in this model, alone or in parallel.
Compared with desktop pump, it has larger amount of exhaust, in particular suitable for big device’s demands.
Technical data:
Model | Power | Voltage | Flow | Body material | Lift | Max. vacuum | Sucking rate for single tap | Quantity of tap | Tank capacity | Dimension | Weight | Security performance |
SHZ-C | 370W | 220V/50Hz | 80L/min | 1cr18 | 12M | 0.098Mpa | 10L/min | 5pcs | 50L | 450×340×870 | 37Kg | check valve |
SHZ-C | 370W | 220V/50Hz | 80L/min | polypropylene | 12M | 0.098Mpa | 10L/min | 3pcs | 50L | 450×340×870 | 37Kg | |
SHZ-95B | 370W | 220V/50Hz | 80L/min | Anti-corrosive | 12M | 0.098Mpa | 10L/min | 5pcs | 50L | 450×340×840 | 37Kg | |
Remark | Equipped with 10L,20L,50L glass reactor or rotary evaporator The vacuum degree is better when the water temperature is between 4℃ and 10℃ |
Zhengzhou Keda Machinery and Instrument Equipment Co.,Ltd.
Our company sells recycled water vacuum pumps, rotary evaporators, thermostatic magnetic stirrers,
glass reactors, low temperature cooling circulating pumps,oil baths, electric mixers, ovens, water baths, freeze-drying machines, electronic energy-saving temperature control devices, smart thermostats and electric experimental equipment sets.
With solid strength, reasonable prices and excellent services, our company has established long-term cooperative relationships with a number of enterprises.
The water pump prices will be packed in wooden cases or as your request.
Shipment:by sea,air or as your desire.
Safe,fast and reliable.
Serial No. | Common faults | Problems may happen | Elimination method |
1 | No vacuum or vacuum vague is not accuracy | 1,Motor speed is not enough or voltage is too low | Check whether the voltage is stable |
2,Vacuum flask, check dirt jam or corrosion | Clean up the internal dirt or replace new bottles and new valve core | ||
3,Water inlet, filter network congestion | Cleaning, take out internal dirt | ||
4,Impeller loss | Replacement impeller | ||
5,Vacuum gauge water or air leakage | Throw water table memory, make the clock back to zero or replace a new table | ||
6,Vacuum line blockage or pipe crack | Clean or replace the new tube | ||
2 | Motor does not work | 1,Insurance tube burn out | Replace the new insurance tube |
2,Capacitance is punctured | Replace the new capacitance voltage is equal to the original capacity | ||
3,Motor winding burn out | Replace the new machine | ||
4,Bearing wear seriously stuck | Replace new bearing | ||
5,The impeller were mountains to death | Clear clutter, correcting impeller position | ||
6,Loose thread virtual welding or line card | Welding, line card is tight | ||
3
| Motor heat, too much noise | 1,Bearing wear | Replace new bearing |
2,The impeller wear deformation | Replace the new impeller | ||
3,Voltage instability or voltage is too low | Check whether the line overload and presence of low voltage phenomenon | ||
4 | Tank water | 1,Filter tube drainage direction is not correct | Adjust the direction of drainage |
2,Water level is too low | Add water to the weir | ||
5 | leaking | 1,Water tank overflow water nozzle loose or seal aging | Tighten the water nozzle or replace the sealing ring |
2,Water tank burst | Replace the water tank |
Products quality guarantee is for one year,after the product was sold.
In the warranty period, the users operate according to instruction and meet problems,
and our company maintains free of charge.
Beyond the warranty period,products problems can also be solved by my company maintenance,
but charge wiring components fees properly.
Hot line:(86)0371-86126825(24H)
Mobile phone:13937197227
Website: /