Immuno Elisa analyzer Automated ELISA Processor Biobase2000
USD $8,000 - $10,000 /Unit
Min.Order:1 Unit
Zhangqiu Meihua International Trading Co., Ltd.
BIOBASE2000 approved by CE,ISO certificate Elisa analyzer,immune analyzer
BIOBASE2000Automated ELISA Processor
BIOBASE2000 Elisa Processors is fully automated Four-plate ELISA processing system,capable of sampling,incubation,washing and data processing.
It is complete solution for your IVD ELISA assays. The system combines a high precision pipetting unit with stable incubators and efficiency washers and photometer unit.
BIOBASE2000 Modularity
1 Robotic hand.
2 Pipetting probe, 1ul stepping (1000ul volumes).
169 Sample rack, 144 Dilution rack.
10 special reagent racks, 2 general reagent racks.
4 units 96 well microplate(pipetting,incubator,reading).
1 unit independent washer&reader( two 8 nozzle manifolds).
Bioelisa Testing Range:
1. TORCH SERIES 4. THYROID AUTOIMMUNITY SERIES
Toxo, RV, CMV, HSV etc. T3, T4, FT3, TSH etc
2. HEPATITIS SERIES 5. ENDOCRINE SERIES
HAV, HBV, HCV, HDV, HEV etc. E2, E3, LH, FSH, HCG, Pro, PRL etc.
3. VENEREAL SERIES 6. TUMOR MARKER SERIES
HIV, TP, MP, CT etc. CEA, AFP, PSA F-PSA, CAL9-9, CAL 25, CA15-3, CA242.
Modularity Composition:
4 microplate ,169 Sample rack, 144 Dilution rack.
All-in-one microplate reader and washer.
10 special reagent racks, 2 general reagent racks.
Robotic precision probes and automated washing system.
Washer:
Two 8 nozzle manifolds
Automatic liquid level sensor detection.
Separate aspiration and dispense probes prevent cross contamination.
Reader:
8 independent photometric channels
Wavelength range 400-700nm.
2 precision filters (450,630nm), 2 more available on request.
User-friendly Software:
LIS system available (laboratory information system)
Large data storage capacity
Professional service software monitor process on line status.
Technical Parameters:
Model | Biobase2000 | |||
Sample pipetting unit | Sample rack | 169 sample tubes(12 mm) | ||
Primary sample tubes | 12*75mm | |||
Pipetting | 8ul~1000ul, (10ul) CV<5% | |||
Time to dispense | 15 minutes (typical)^100ul of 96 reagents to plate from sample tubes or dilution tubes. | |||
Dispensing system | 1 aspirating-dispensing probe with X-Y-Z movement;dilution system with a syringe of 1000ul, continuous flow washing station. | |||
Reagent pipetting unit | Reagent racks | 10 special reagent racks,2 general reagent racks | ||
Removable trays | modular racks to hold calibrators, controls and reagents for 10 methods on line | |||
Pipetting | 8ul~1000ul,1ul stepping, | |||
pipetting precision | (100ul )CV≤0.2~1% | |||
Time to dispense | 2 minutes(typical)^100ul of 96 reagents to plate from reagent bottle | |||
Washing unit | Programmable | Yes(options & parameters) | ||
Wash containers | 4 wash buffers at 2.0L, with level-sensing | |||
Waste containers | 10L with waste full sensor | |||
Washing system | two 8 nozzle manifolds, one to dispense and the other one to aspirate; | |||
Washing residual | <2µl | |||
Reading unit | Reading system | 8 independent photometric channels with mono and bichromatic reading capability | ||
Photometric range | 0.000 ~3.000 OD | |||
Spectral range | 400-700nm | |||
Optical filters | 2 precision filters (450and 630nm); 2 more on request (4 filters max, 400-700nm) | |||
Reading inaccuracy | inaccuracy±1% from 0.000 to 1.500 OD, ±2% from 1.500 to 3.000 OD | |||
Incubation unit | Incubators | 4 independent units | ||
Programmable | Yes | |||
Temperature range | RT to 45 °C | |||
Temperature accuracy | ±0.2°C | |||
Adjustment step size | ±0.1°C | |||
Software unit | Software system | Windows XP or Windows 7 or above, LIS system available | ||
Function application | work list set up, patient reports and result filing; calibration curve storage for each carried out test | |||
Data reduction | cut-off (qualitative) | |||
Physical Specification | Dimensions | 1130x750x770mm (W*D*H) | Weight | 120kg |
Power Supply | AC110/220V+ 10%,50/60Hz | Consumption | 350W |