BIOBASE 4000 2 probes, 4 microplates Fully Automatic ELISA Processor, Elisa Analyzer, Elisa Kits
USD $15,000 - $15,500 /Unit
Min.Order:1 Unit
Zhangqiu Meihua International Trading Co., Ltd.
BIOBASE 4000 2 probes, 4 microplates Fully Automatic ELISA Processor, Elisa Analyzer, Elisa Kits
1 Robotic hand.
4 Pipetting probe, 1ul stepping (1000ul volumes).
Total 25 racks, arbitrary rack can be defined as sample or reagent on the software.
4 units 96 well microplates (pipetting,incubator,reading).
1 unit independent washer &reader( two 8 nozzle manifolds).
Testing Range:
1. TORCH SERIES 4. THYROID AUTOIMMUNITY SERIES
Toxo, RV, CMV, HSV etc. T3, T4, FT3, TSH etc
2. HEPATITIS SERIES 5. ENDOCRINE SERIES
HAV, HBV, HCV, HDV, HEV etc. E2, E3, LH, FSH, HCG, Pro, PRL etc.
3. VENEREAL SERIES 6. TUMOR MARKER SERIES
HIV, TP, MP, CT etc. CEA, AFP, PSA F-PSA, CAL9-9, CAL 25, CA15-3, CA242.
Model | BIOBASE4000 | |
Sample pipetting unit | Sample rack | 24 sample tubes/rack, at most 25 racks |
Primary sample tubes | 11.5-13mm | |
Pipetting | 8ul~1000ul, (10ul) CV<5% | |
Time to dispense | 8 minutes (typical)^100ul of 96 samples to plate from sample tubes or dilution tubes. | |
Dispensing system | 1 aspirating-dispensing probe with X-Y-Z movement;dilution system with a syringe of 1000ul, continuous flow washing station. | |
Reagent pipetting unit | Reagent racks | 2 reagent tanks(60ml)/ rack, at most 25 racks |
QC racks | 24 positions/QC rack, at most 25 racks | |
Pipetting | 8ul~1000ul,1ul stepping, | |
pipetting precision | (100ul )CV≤0.5% | |
Time to dispense | 2 minutes(typical)^100ul of 96 reagents to plate from reagent bottle | |
Washing unit | Programmable | Yes(options & parameters) |
Wash containers | 4 wash buffers at 2.0L, with level-sensing | |
Waste containers | 10L with waste full sensor | |
Washing system | two 8 nozzle manifolds, one to dispense and the other one to aspirate; | |
Washing residual | <2μl | |
Reading unit | Reading system | 8 independent photometric channels with mono and bichromatic reading capability |
Photometric range | 0.000 ~3.000 OD | |
Spectral range | 400-700nm | |
Optical filters | 2 precision filters (450and 630nm); 2 more on request (4 filters max, 405 nm, 450 nm, 492 nm and 630 nm) | |
Reading inaccuracy | inaccuracy±1% from 0.000 to 1.500 OD, ±2% from 1.500 to 3.000 OD | |
Incubation unit | Incubators | 4 independent units |
Programmable | Yes | |
Temperature range | RT to 45 °C | |
Temperature accuracy | ±0.5℃ | |
Temperature staility | ±0.5℃ | |
Software unit | Software system | Windows XP or Windows 7 or above, LIS system available, realize two-way data communications |
Function application | work list set up, patient reports and result filing; calibration curve ,storage for each carried out test | |
Data reduction | cut-off (qualitative) | |
Physical Specification | Power Supply | AC220V+ 10%,50/60Hz,450W |
Instrument Dimension | 110*72*70cm (L*W*H) | |
Instrument Weight | 100kg | |
Package Dimension | 138*91*98cm(L*W*H) | |
Total Weight | 150kg |