XINFU Specific Heat Calorimetry
Negotiable
Min.Order:1
Zhengzhou Prominent Technology Co., Ltd.
GC106 Calorimeter is new developed model based on ZDHW-5 Calorimeter.
General Introduction of Specific Heat Calorimetry
Calorimeter calorimeter system, in addition to the water absorbs heat, the oxygen bomb, inner tube, thermometer and stirrer, and this will absorb heat, and absorbing their situation is different, more complex variety of factors, it is impossible to rely on a simple mathematical calculations to obtain, but to use the reference material of known calorific value such as benzoic acid to the actual calibration of the system temperature is increased by the amount of heat to be absorbed heat 1 ℃, which is calibrated out of the heat capacity of the calorimeter.
Under High Pressure Oxygen Condition:
Put samples in hot-corrosion-resistant metal crucible.
Put this crucible in oxygen bomb and seal it.
Charge oxygen in the bomb until the pressure reach 3.0Mpa
Put oxygen bomb in and turn on the calorimeter, burn the sample.
The heat will pass to the water in cylinder.
Calorimeter can calculate the calorific value of samples according to measure the water temperature, cylinder and bomb.
Features of GC106
1, the use of advanced single-chip system, the instrument has automatic water, drainage, auto temperature, automatic mixing, automatic ignition, automatic printing and other functions, simple operation, long continuous measurement. 2, large-screen LCD display, Chinese menu prompt, simple and easy to operate. 3, with the experiment and converted high calorific functions. 4, automatic cooling calibration experiments, the use of ambient temperature requirements relaxed. 5, the automatic tracking the ambient temperature, the water temperature need not be measured outside the barrel.
Working Principle of Specific Heat Calorimetry
Stirrer is composed of mini electric motor, stir vane and stirring shaft. The stirring speed is stable. Inner cylinder frame is made of plexiglass or bakelite. Oxygen charger is special designed to work with calorimeter, it can completely avoid splash during the charging. The charging speed can be adjusted. It is very easy to operate.
Standard Reference
ASTM D3286-96: Standard Test Method for Gross Calorific Value of Coal and Coke by the Isoperibol Bomb Calorimeter (Withdrawn 2000) ASTM D1989-97: Standard Test Method for Gross Calorific Value of Coal and Coke by Microprocessor Controlled Isoperibol Calorimeters (Withdrawn 2000) ASTM D5865-13: Standard Test Method for Gross Calorific Value of Coal and Coke ASTM D2015-85: Standard Test Method for Gross Calorific Value of Coal and Coke by the Adiabatic Bomb Calorimeter
Main Specification