V-1600 Lab Visible Spectrophotometer
USD $800 - $1,300 /Unit
Min.Order:1 Unit
Hangzhou Chincan Trading Co., Ltd.
V-1600 SPECTROPHOTOMETER
V-1600 spectrophotometer is simple-to-use instruments with advanced performance; its stray light is only 0.05%T. The local stand-alone software provides functions of Photometry, Quantitative Test, Kinetics and System Utilities functions.
Specifications
Model | V-1600 |
Wavelength Range | 320-1100nm |
Spectral Bandwidth | 4nm |
Optical System | Single Beam,Grating 1200 lines/mm |
Wavelength Accuracy | ±0.5nm |
Wavelength Repeatability | 0.3nm |
Photometric Accuray | ±0.5%T or ±0.004A@1A |
Photometric Range | 0-200%T,-0.3-3A,0-9999Conc |
Stray Light | 0.05%T@360nm |
Stability | ±0.001A/h@500nm |
Display | Graphic LCD (128*64 dots) |
Keyboard | 22 membrane keypad |
Standard Cell Holder | 4-position 10mm cell changer |
Sample Compartment | Standard 10mm pathlength cuvette |
Light Source | Tungsten lamp |
Output | USB Port & Parallel Port (Printer) |
Power Requirement | AC 110/220V 50/60HZ |
Dimension(L*W*H) | 470*373*187mm |
Weight | 12kg |
Features
1. Large LCD screen (128*64 Dots).
2. System can also save the test results, total 200 groups of data 200 standard curves can be saved in the RAM. Convenient for check and reload.
3. Data can be restored after a sudden power cut.
4. Auto setting wavelength.
5. Tungsten & Deuterium lamp can be turned on/off individually to extend lifetime.
6. The optional application software M. Wave Professional provides complete control of the spectrophotometer from a computer through the Built-in USB port. It can expand to the following functions: Quantitative, Kinetics, Wavelength Scan, Multi-wavelength & DNA/Protein Test.
7. Pre-aligned design makes it convenient to change lamps.
8. Large sample compartment, it can accommodate 5-100mm path length cuvettes with optional holders. A variety of optional accessories are available.
Functions
1. Photometry
Absorbance, Transmittance or Energy measurements. It can display and save 200 groups of data, 5 groups per screen.
2. Quantitative.
2.1 Standard Curve
At most, 9 standard samples can be used to establish a standard curve. The curve and its equation will be displayed on the screen simultaneously. You can measure your unknown concentration solutions by the curve,total 200 curves can be saved in the Memory.
2.2 Coefficient Method
If the coefficient k & b in the formula C=k*a + b is known, you can input them directly by the button, and then test your unknown solutions.
3. Kinetics.
This mode may be used for time course scanning or reaction rate calculations. Abs. VS Time graphs is displayed on the screen in real time. It can most record 1000 data.
4. System Utilities
Lamp management, time & date set, obtain dark current, calibrate wavelength, default system, and some other system functions.
Brief Function Introduction
Main Menu
Move the cursor on the function menu you want, then press ENTER key to go into the corresponding interface.
Basic Mode
Absorbance, Transmittance or Concentration measurements. It can display and save 200 groups of data, 5 groups per screen.
Quantitative
2.1 Standard Curve
At most 9 standard samples can be used to establish a standard curve. The curve and its equation will be displayed on the screen simultaneously. You can measure your unknown concentration solutions by the curve. Total 200 curves can be saved in the RAM.
2.2 Coefficient Method
If you have known the coefficient K*a + b is known, you can input them directly by the button, and then test your unknown solutions.
Kinetics
This mode may be used for time course scanning or reaction rate calculations. Abs. VS Time graphs is displayed on the screen in real time. It can most record 1000 data.
System Utilities
Lamp management, time & date set, obtain dark current, calibrate wavelength, default system, and some other system functions.