neutral electrode contact impedance tester
Negotiable /Set
Min.Order:1 Set
HFPA150
The simple and convenient neutral electrode contact impedance tester.
Standard Coverage: IEC 60601-2-2
OVERVIEW
Designed according to IEC 60601-2-2:2009 clause 201.15.101.6 for testing NE contact impedance
Outputs 200mA into 0-50Ω load from 50KHz to 5MHz
Equipped with DDS (direct digital synthesis) technology for a stable frequency output
Inbuilt current monitor port eliminating the need for expensive current probes
SPECIFICATION
Parameters Specifications
Frequency Range 50kHz – 5MHz
Maximum Voltage 11Vrms (~31Vpp), ≥ 50Ω
Maximum Current >200mArms, ≤ 50Ω
Waveform Sine
Frequency Response <±0.8dB, 50Ω
THD<0.3% (-50.4dB), at 200mArms, 50Ω, 1MHz
Impedance<3.2Ω + j0.4µH
Auxiliary Equipment for Contact Impedance Test Oscilloscopes (Ch1 connects with voltage probe to measure voltage, Ch2 connects with HFPA150 current monitoring interface to measure the current value)
Current Measurement Accuracy <±5%
Current Measurement Frequency Range 50kHz – 5MHz
Current Measurement Connector Type BNC
Power Source
Power Adapter
Input: 100 – 240V, 50/60Hz
Output: 12V DC / 2A
Environment 10℃ – 40℃, 30% – 90% RH
Size / Weight 18 x 18 x 4.8cm (L x W x H) / 0.7kg (net weight)