SECG 5.0 AIO
The ECG simulator designed for ECG performance testing. Capable of adding noise, respiration and lead-off simulations. Flexible operation modes for standalone or PC software.
Suitable for R&D and compliance testing.
Compliant with IEC 60601-2-25, IEC 60601-2-27, IEC 60601-2-47, GB 9706.225, GB 9706.227, YY 9706.247, YY1079, YY1139, YY0782, YY0885, JJG 760, JJG 1042, ANSI/AAMI EC11, ANSI/AAMI EC13
SPECIFICATION
ECG Test Mode
Parameters Specifications
Main Output Voltage Accuracy ±1% for amplitudes of 0.5mVpp or higher
Main Output Voltage Resolution
(DAC resolution) 0.63µV
Frequency / Pulse Repetition Rate Accuracy ±0.5%
Pulse Duration / Timing Accuracy (excluding pacing) ±0.5ms
ECG Duration Accuracy ±0.5ms
Pacing Pulse Width Accuracy ±5μs
Pacing Pulse Amplitude Accuracy, Range ±2mV pulse: ±0.3%
>2mV pulse: ±10%
Range: ±2mV ~ ±700mV
Pacing Pulse Characteristics Rise/fall time 5μs
Overshoot <1%
Settling time <1%
Pacing Pulse Overshoot (intentional) Method A according to IEC 60601-2-27
Resistor Tolerance ±1%
Capacitor Tolerance ±5%
Precision 250:1 Divider ±0.1%
Sample Rate 10kHz ± 0.05%
DC Offset
(fixed, noise-free, sourced from the internal supercapacitor) 300mV ± 1%
DC Offset
(variable, up to 1000mV, may include up to 50μVpp noise) ±1000mV ±1% or ±3mV
Environment 5 ~ 40°C
50 ~ 80% RH
altitude <2000M
Overvoltage CAT II
Signal Type
Parameters Setting Range Default Values Minimum Step Size
Sine Frequency (Hz) 0.05 ~ 500Hz 1Hz 0.01
Frequency (BPM) 3 ~ 30000BPM 60BPM 1
Amplitude (-40) ~ 40mV 1mV 0.01
Triangle Frequency (Hz) 0.05 ~ 500Hz 1Hz 0.01
Frequency (BPM) 3 ~ 30000BPM 60BPM 1
Amplitude (-40) ~ 40mV 1mV 0.01
Square Frequency (Hz) 0.05 ~ 500Hz 1Hz 0.01
Frequency (BPM) 3 ~ 30000BPM 60BPM 1
Amplitude (-40) ~ 40mV 1mV 0.01
Rectangle Pulse Frequency (Hz) 0.05 ~ 5.00Hz 1Hz 0.01
Frequency (BPM) 3 ~ 300BPM 60BPM 1
Amplitude (-20) ~ 20mV 1mV 0.01
Pulse Width 1 ~ 300ms 100ms 1
Triangle Pulse Frequency (Hz) 0.05 ~ 5.00Hz 1Hz 0.01
Frequency (BPM) 3 ~ 300BPM 60BPM 1
Amplitude (-20) ~ 20mV 1mV 0.01
Pulse Width 1 ~ 300ms 100ms 1
Exponential Frequency (Hz) 0.05 ~ 5.00Hz 1Hz 0.01
Frequency (BPM) 3 ~ 300BPM 60BPM 1
Amplitude (-20) ~ 20mV 1mV 0.01
ECG Waveform P Amplitude 0 ~ 2mV 0.2mV 0.01
P Duration 50 ~ 100ms 100ms 1
QRS Amplitude (-20) ~ 20mV 1mV 0.1
QRS Duration 5 ~ 200ms 100ms 1
PR Interval 0 ~ 200ms 160ms 1
QT Interval 100 ~ 900ms 350ms 1
ST Deviation (-2) ~ 2mV 0mV 0.01
T Amplitude 0 ~ 5mV 0.2mV 0.01
Frequency (Hz) 0.02 ~ 6Hz 1Hz 0.01
Frequency (BPM) 1 ~ 360BPM 60BPM 1
Amplitude (-20) ~ 20mV 1mV 0.1
Signal Add-on
Parameters Setting Range Default Values Minimum Step Size
DC Offset (-1000) ~ 1000mV 0mV 1
Pacing Amplitude (-700) ~ 700mV 0mV 1
Duration 0.1 ~ 3ms 2ms 0.1
Pacing Rate 10 ~ 360BPM / Synchronization 10BPM 1
Overshooting (Amplitude) 0 ~ 5mV / Amplitude automatic 0mV 0.01
Overshooting
(Time Constant) 0 ~ 100ms 0ms 1
Baseline Overload Test Frequency 50 / 60Hz 50Hz -
620kΩ / 4.7 nF on / off Off -
Sine Wave Add-on on / off Off -
Noise Main Noise 50Hz / 60Hz / White 50Hz -
Amplitude 0.01 ~ 5mV 0.1mV 0.01