Acrel AID150 medical digital centralized alarm insulation display instrument indicator device for hospital operating room
Negotiable /Piece
Min.Order:1 Piece
AID series alarm and display device is a remote display and sound and light alarm device by using advanced fieldbus communications technology. The device can be installed on the operating room or ICU, the medical staff can easily know the status of medical isolated power systems, and the type of fault when the system fails.
-The insulation resistance alarm value, load current alarm value and transformer temperature alarm value of the system can be set remotely. -When the system appears insulation fault, overload, transformer temperature over limit and wiring fault, alarm and display instrument gives out the corresponding sound and light alarm, and has the function of eliminating sound alarm. -Using advanced fieldbus technology, the remote monitoring function can be realized by real-time data interaction with the insulation monitor.
Parameter/Model | AID10 | AID120 | AID150 | AID200 | |
Auxiliary power | Voltage | DC 24V | |||
Power consumption | <0.6W | ||||
Insulation resistance measurement range | 0-999kΩ | / | |||
Insulation alarm value range | 50-999kΩ | / | |||
Transformer load rate display | Percentage display | / | |||
Load current alarm settings | 14A、18A、22A、28A、35A、45A | ||||
Temperature alarm value range | 0~+200℃ | ||||
Alarm | Sound and light alarm | ||||
Alarm type | Insulation fault, overload, over-temperature, equipment failure | ||||
Communication | RS485 interface, MODBUS-RTU protocol | CAN interface, a custom protocol | |||
Display method | LED display | 128*32 LCD |