BIOBASE China immunoassay analyser 120T/H clinic laboratory Automatic Chemiluminescence analyzer BKI2200
Negotiable
Min.Order:1
Sample Tray | Reagent Tray | Washing and mixing module | Sample Probe |
1. Disc design, up to 60 samples loading capacity per batch 2. The position of sample, standard product and quality control positions can be edited 3. Support emergency function. | 1. Patented semiconductor refrigeration technology 2. Up to 30 reagent position on board with refrigeration at 2-8℃ 3. Up to 30 items can be tested at the same time. | 1. Three-level cleaning mode, higher cleaning efficiency; 2. Non-contact mixing to avoid cross-contamination. | 1. Excellent rinsing system make carry-over ≤10-5 2. Liquid-level and blood coagulation detection system 3. Intelligent anti-collision function 4. Sampling volume: 15μl-100μl |
RV Loading Module | Incubation plate | Instrument Software | |
1. Independent reaction vessel 2. Up to 1200 pcs per batch with continues loading 3. Automatically supply reaction vessel 4. Intelligently detect the amount of RV | 1. 80 incubation positions 2. Constant 37°C incubation | 1. Flexible sample test mode: STAT function, automatic dilution, reset function, continuous loading 2. Reagent management: the remaining amount, calibration validity, expiry date management 3. Consumables management: continuous monitoring and low-level warning 4. Result management: store search, print and statistics, automatic unit conversion function | LIS/HIS connection Auto daily maintenance |
Model | BKI1100 | BKI2200 | |
Throughput | 80T/H | 120T/H | |
Principle | Magnetic particle enzymatic chemiluminescence | ||
Modes of Operation | Random, Batch and STAT | ||
Separation method | Magnetic separation technology | ||
Sample Positions | 60 (Each position could be used as an emergency position) | ||
Reagent Positions | 30 pcs (2-8℃ refrigeration) | ||
Incubation Positions | 50 | 80 | |
Time to 1st Result | 18 Minutes | ||
Sample Volume | 15-100μl | ||
Detection Precision | CV≤5% | ||
Carry-over Rate | ≤10-5 | ||
Correlation Coefficient | r≥0.99 | ||
Calibration Stability | 28 Days | ||
Calibration Type | 6/7-point calibration | ||
Software system | Windows 7/8/10, 32 or 64 bit Bi-direction, support HL7 protocol Intuitive User Interface, Intelligent Data Management | ||
Interface | TCP/IP Network interface | ||
Other Function | Liquid level detection Anti-collision function Sample barcode scanning Reagent radio frequency identification Intelligent alarm prompt | ||
Power Supply | AC110/220V,50/60Hz | ||
External Size (W*D*H) | 1120*735*513mm | 1305*770*1150mm | |
Net Weight | 130Kg | 220Kg | |
Package Size (WD*H) | 1303*873*854mm | 1608*898*1534mm | |
Gross Weight | 180kg | 294kg |