Model | BIOBASE2000 | ||
Sample pipetting unit | Sample Racks | 552 sample positions (tubes 13mm) | |
Primary sample tubes | 11.5-13mm | ||
Pipetting | 8μl~1000μl, (10μl) CV<5% | ||
Time to dispense | 12 minutes adding 96 samples | ||
Dispensing system | 1 aspirating-dispensing probe with X-Y-Z movement;dilution system with a syringe of 1000μl, continuous flow washing station. | ||
Reagent pipetting unit | Reagent racks | 2 reagent tanks(60ml)/ rack, at most 23 racks | |
QC racks | 24 positions/QC rack, at most 23 racks | ||
Pipetting | 8μl~1000μl,1 μl stepping | ||
pipetting precision | (100μl )CV≤0.5% | ||
Time to dispense | 3 minutes(typical)^100μl of 96 reagents to plate from reagent bottle | ||
Washing unit | Programmable | Yes(options & parameters) | |
Wash containers | 3 wash buffers at 2.0L,1 wash buffers at 5.0L,with level-sensing | ||
Waste containers | 25L with waste full sensor | ||
Washing system | two 8 nozzle manifolds, one to dispense and the other one to aspirate | ||
Washing residual | <2μl | ||
Reading unit | Reading system | 8 independent photometric channels with mono and bichromatic reading capability | |
Photometric range | 0.000 ~3.000 OD | ||
Spectral range | 400-700nm | ||
Optical filters | 2 precision filters (450and 630nm); 2 more on request (4 filters max, 405 nm, 450 nm, 492 nm and 630 nm) | ||
Reading inaccuracy | inaccuracy±1% from 0.000 to 1.500 OD, ±2% from 1.500 to 3.000 OD | ||
Incubation unit | Incubators | 4 independent units | |
Programmable | Yes | ||
Temperature range | RT to 45 °C | ||
Temperature accuracy | ±0.5℃ | ||
Temperature staility | ±0.5℃ | ||
Software unit | Software system | Windows 7/8/107 or above, LIS system available | |
Function application | work list set up, patient reports and result filing; calibration curve storage for each carried out test | ||
Data reduction | cut-off (qualitative) | ||
Others | Power Supply | AC 110/220V±10%,50/60Hz,450W | |
External Size(W*D*H) | 1235*710*880mm | ||
PackageSize(W*D*H) | 1380*860*1435mm | ||
Gross Weight | 220kg |
Air protection product | Cold chain product | General lab equipment | Analytical instrument |
biosafety cabinet | medical refrigerator | autoclave | Chemistry Analyzer |
Laminar flow cabinet | blood bank refrigerator | incubator | ELISA Reader& Washer |
fume hood | -25 freezer | drying oven | Hematology Analyzer |
clean room | -40 freezer | centrifuge | Electrolytic analyzer |
PCR cabinet | -60 freezer | spectrophotometer | Urine Analyzer |
ect | -86 freezer | shaker | ect |