BIOBASE China Auto Hematology Analyzer BK-6190 Hematology Analyzer 3-Part Blood Testing CBC Machine for lab
Negotiable
Min.Order:1
Model | BK-6190 | ||||||
Throughput | 60Tests/hour | ||||||
Measuring Principle | Impedance for cell counting Cyanide-free method for HGB | ||||||
Aperture Diameter | WBC chamber:100μm,RBC/PLT chamber:70μm | ||||||
Parameters | WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%, RBC, HGB, HCT, | ||||||
MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR | |||||||
Histograms | 3 histograms(includingWBC, RBC, PLT histogram) | ||||||
Sample Volume | Normal:10μl; Predileted:20μl | ||||||
Calibration | Manual & auto calibration | ||||||
Counting Modes | Venous, capilary, prediluted | ||||||
Printer | Built-in thermal printer,USB to external printer(optional) | ||||||
Data Input | 8 inch touch screen, keyboard, mouse | ||||||
Output | RS232 port, USB port, VGA port. LIS system avaliable | ||||||
Storage | 100,000 results with histogram | ||||||
Carry-over rate | Parameter | CV | |||||
WBC | ≤ 2% | ||||||
RBC | ≤ 1% | ||||||
HGB | ≤ 2% | ||||||
PLT | ≤ 2% | ||||||
Background | Parameter | Background | |||||
WBC | ≤ 0.2×10^9/L | ||||||
RBC | ≤ 0.02×10^12/L | ||||||
HGB | ≤ 1 g/L | ||||||
HCT | ≤ 0.5% | ||||||
PLT | ≤ 10×10^9/L | ||||||
Precision | Parameter | CV | Measurement range | ||||
WBC | ≤ 2.0% | (4.0-15.0×10^9)/L | |||||
RBC | ≤ 1.9% | (3.5-6.0×10^12)/L | |||||
HGB | ≤ 1.9% | (110.0-180.0g)/L | |||||
MCV | ≤ 0.4% | (80.0-110.0)fL | |||||
PLT | ≤ 4.0% | (100.0-500.0×10^9)/L | |||||
Linearity | Parameter | Measurement range | CV | ||||
WBC | (0-5.9×10^9)/L | (±0.3×10^9)/L | |||||
(6.0-99.9×10^9)/L | ±5% | ||||||
RBC | (0-0.99×10^12)/L | (±0.05×10^12)/L | |||||
(1.0-9.99×10^12)/L | ±5% | ||||||
HGB | (0-99.0g)/L | (±2.0g)/L | |||||
(100.0-300.0g)/L | ±2% | ||||||
PLT | (0-99×10^9)/L | (±8.0×10^9)/L | |||||
100-999×10^9)/L | ±10% | ||||||
Power supply | AC 220±10%,60/50Hz, 110±10%,60Hz | ||||||
Package Size | 580*470*590mm (W*D*H) | ReagentPackage Size | 450*410*310mm (W*D*H) | ||||
Gross Weight | 27kg | ReagentGross Weight | 23kg |