1 Flat Panel Detector
1.1 Amorphous silicon flat panel detector
1.2 Effective pixels:≥9 million
1.3 Pixel matrix:≥3070*3070 pixels
1.4 Quantization depth:≥16bit
1.5 Effective area:17”*17”
1.6 Spatial resolution:≥3.7Lp/mm
1.7 Imaging time:≤7s
2 Mechanical structure: scythe arm(type Z)
2.1 Vertical range(scythe arm):600mm ~ 1600mm(±5%)
2.2 Rotation range of the scythe arm:-45°~ +135°(±2°)
2.3 SID (vertical and horizontal)can be adjusted electronically,the range:1000mm~1800mm(±5%)
2.4 Rotation range of the tube assembly :-15°~ +15°(±2°)
2.5 Rotation range of the detector :-45°~ +45°(±2°)
2.6 The detector end can independently control the movement of the stand
2.7 9.5 "LCD touch screen
2.8 Minimum height of the equipment room:2.4m
2.9 The tube and detector move synchronously to maintain the balance of the stand
3 High Voltage Generator
3.1 High voltage generator smart network voltage real-time monitoring system(smart monitor)
3.2 Power:≥50kW
3.3 kV range:40kV-150kV
3.4 mA range:10mA-630mA
3.5 mAs range:0.1mAs-630mAs
3.6 Maximum operating frequency:200kHz
4 X-ray Tube
4.1 Focus size:≤0.6mm/1.2mm
4.2 Focus power:
Small focus:≥ 20kW
Large focus:≥ 50kW
4.3 Maximum heat capacity:≥300kHU
4.4 Maximum peak voltage:≥150kV
5 Filter grid
5.1 Ratio:≥10:1
5.2 Density:≥40L/CM
6 LED Collimator
6.1 Time limit of indicator :30±3s
6.2 Inherent filtration:1.0 mm Al.EQ.
7 Workstation
7.1 DELL 5040MT,Intel Pentium G4400 processor, 4G/500G/DVDRW/WIN10,E2216H 21.5 inch high contrast display
8 Software function(patient management)
8.1 Remote direct access patient information in RIS with worklist protocol;
Manually create the patient examination information;
Emergency quick registration;
Simple,advanced, or custom data query methods;
The image CD burning in Dicom standard ,can be viewed on any standardimage workstation;
Inquiry and management of historical image data;
Detection disk space, automatic cleaning of old check data;
Dicom transmission of image and seamless connection with PACS;
Support diagram text report,carry report library;
Print patient's reports, exposure times can be displayed in the list of patients;
The original image is easy to view.
9 Software function(software direct exposure control)
9.1 Support customization for dose template;
Selection part and position template, automatically bring out the matching dose;
Selection shape of the patient, automatically select the matching dose, and adjust to the dose online;
Control the size of focus;
Display generator's status transparent;
Refusal and acceptance of images;
Support automatic sending of images.
10 Software function(image processing)
10.1 Supports dual or multi-screen images display;
Manually/automatically/preset window width and window level, partial window width and window level;
Operation positive and negative of image,flip image,rotation image,scaling image and roaming;
Automatic or manual addition of image information;
Line, angle, rectangle, ellipse, polygon and other measuring tools;
Balance tissue,enhancement contrast,optimization dose;
Enhancement edge: automatic recognition and analysis of images,enhance edge sharpness.
10.2 Preview the post-processing image 1:1: the doctors make more intuitive diagnosis in image optimization;
CAD View: just click the mouse,doctors can see the details of the image more clearly, so that doctors can make more accurate diagnosis.
11 Software function(film printing)
11.1 Set film properties, image layout, printing mode;
Typeset quickly by manual/automatic ;
Select any camera within the network;
Customization of patient information and display location;
Support for queue management;
Support for setting print priority