It is suitable to determine flash point and fire point of petroleum products having a flash point higher than 79 °C, excepting fuel oils.
Standard: ISO2592, ASTMD92, IP36
specification
1. Power supply: AC 220V±10%, 50 Hz;
2. Heating device: It is a quartz tube heater, so there is no flame appears. Explosion protective; its power can be continuously adjusted in the range of 0∼600W;
3. Flame application device: automatically scan;
4. Thermometer: -6~400°C; Each scale division is 2 °C;
5. Ignition device:
(1) Ignition source: coal gas (or any other combustible gas); Automatic
(2) Spout aperture: 0.6∼0.8 mm;
6. Ambient temperature: -10°C∼50°C;
7. Relative humidity: ≤85%;
8. Total power consumption: not more than 400 W
9. Dimension: 35*30*40cm
10. Weight: 10kg