DW-TEK8510 Drawell auto 5 part Hematology Analyzer price (27Parameter + 4Histogram)
USD $7,150 - $10,000 /Piece
Min.Order:1 Piece
Shanghai Drawell Scientific Instrument Co., Ltd.
DW-TEK8510 5-Automatic Hematology Analyzer for Clinic
A real automatic 5 diff hematology analyzer but only
l Small space:53cmx40cmx52cm
l Easy to carry: weight 28kg
Opportunity to replace 3-diff hematology analyzer
l 5 diff, 27 parameters
l 1 Scattergrams, 4 Histograms
l Up to 60 samples per hour
l Optical solid state LASER scattering
l High efficient light collector (patented)
l 12.1" TFT Touch Screen
l Original reagent, control and calibration
l Easy to operate: Friendly and Intellective Software
Automatic Hematology Analyzer
Technical Specifications
Principle
Optical solid state LASER scattering Wide angle light scatter,High efficient light collector(patented)
Electrical Impedance,Double hydrodynamic focusing
Parameters
27 parameters:WBC, LYM#, M0NO#, GRA#, LYM%, M0NO%, GRA%, RBC, HGB,
HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PDW, PCT and Histogram for WBC, RBC, PLT
1 Scattergrams:WBC Interpreted Scattergram
4 Histograms:RBC,PLT,WBC,cyto
Throughput
Up to 60 samples per hour
Display
12.1Inch Touch Screen
Sample types
Whole blood,PLT only,HB only,Capillary predilution,Diluent,Control,Calibrator
Performance
Parameter | Linearity range | Precision (CV%) |
WBC | 0.1-450 | 2.5(7.0-15.0) |
RBC | 0.2-8.5 | 2.5(3.50-6.0) |
HGB | 2-250 | 2.0(110-180) |
MCV | 45-135 | 0.5(80.0-110.0) |
PLT | 10-4000 | 5.0(200-500) |
Sample volume
Prediluted: 20µL
Whole blood: 10µL
Carryover
WBC<0.5%, RBC<0.5%, HGB<0.5%, PLT<1.5%
Mult-language
Chinese,English for optional and others can be customized
Data storage capacity
Up to 100,000 results including numeric and graphical information
Printout
External Laser printer/Inkjet printer, various printout formats and user-defined formats
Operating Environment
Temperature: 18-32℃ Humidity: 30%-85%
Power Requirement
AC 100-240V 50/60HZ
Dimension
Instrument:532mmx400mmx520mm
Package: 696mmx576mmx696mm
Weight
Net:28KG Gross:32KG
Laser Light scatter and Cell Histochemistry method
hematology analyzer utilizes the reliable and economical semiconductor laser as the light source for the cytometry system. Through analysis of different angles of laser scatters, the instrument provides complete analysis including cell size,granularity and complexity.
Accurate blood dispense system
utilizes advance ceramic sample rotator valve platform to accurately segments blood with high precision.
Original Reagent
Only 3 types:Sheath Hematon,Enzyme Detergent and Complex Cell Hemalyse
High quality original reagent ensure high accuracy.
More than one year expiration time.
Superior ability to flag abnormal cells
Utilizing laser scatter technology and histochemistry method, TKE8510 can identify abnormal and suspect information and flag the abnormal cells to help decide if the sample needs recount.
Convenient sample results review
200,000 sample results with numeric and graphical information can be searched,monitored, analyzed and printed conveniently.Eight methods to review the test results are provided for the operator’s option.
Dual Sheath Flow Cell System and Intelligent Sample Treatment System
Dual sheath flow system reduces the velocity gap between sheath flow and blood cells,ensuring a more stable signal.Through the use of an intelligent sample treatment and hydrokinetics,single blood cell flow is facilitated ensuring accurate measurement.
Pre-heater