Wireless temperature / humidity / pressure verification system
USD $3000 - $6000 /Set
Min.Order:1 Set
Tai'an Dearto Automation Instrument Co.,Ltd
Outstanding performance oversized thermometer and hygrometer calibration chamber
Programmable Constant Temperature And Humidity Test Chamber for hygrothermograph calibration
High Precision constant temperature cabinet for testing thermometer
Automatic Multi-furnace group control system for thermocouple calibration
0.001 deg C resolution portable dry block well -30~150 C low temperature calibrator
Portable dry block well type 50 to 1200 deg C temperature calibration furnace
DT Pro wireless temperature / humidity / pressure verification system can be widely used in equipment verification and environmental monitoring in pharmaceutical, food, biological, chemical and other industries. Compared with the traditional wired verification system, the probe can be conveniently and quickly arranged, and the detection point is not physical constraints greatly improve verification efficiency.
Function
The system consists of a wireless recorder (temperature,humidity, pressure), a data reading workstation, and wireless verification software.
The temperature recorder can effectively store 16000 pieces of data, and the temperature measurement accuracy is 0.1 ℃, which fully meets the needs of equipment and process verification in food and drug production enterprises and the verification needs in the fields of bioengineering, medical and transportation. The recorder uses replaceable batteries, with large power, low power consumption and long battery life. Adopt high pressure and waterproof design, can be used continuously for a long time in high pressure and high humidity environment.
The data reading workstation is divided into a single-hole workstation and a 10-hole workstation. The universal USB serial port is used to connect to the computer, and data communication can be carried out conveniently, efficiently and safely.
The software realizes data collection, data analysis and report generation, meets FDA 21 CFR Part 11, and supports electronic signature of documents. Support custom stages and verify each stage to generate stage verification report.
Features:
- Switch between Chinese and English versions at will;
- Comply with FDA 21CFRPart 11 regulations;
- With data audit tracking function, with three levels of authority (freely assignable authority);
- It has data statistical analysis functions, including large and small values, average value, span, temperature difference at the same time, F0 value test, etc., and can be freely selected according to needs.
Application scenarios:
The system can be applied to GMP pharmaceutical industry, GSP industry, biological industry, medical industry, food industry, transportation industry.
GMP | Steam sterilizer verification / water bath sterilizer verification, rubber plug washer verification / aluminum lid washer verification, freeze dryer verification / cryogenic refrigerator verification, constant temperature and humidity chamber verification / stability incubator verification |
GSP | Cold storage verification / refrigerated truck verification, refrigerator / incubator verification, cool storage / normal temperature storage verification |
Biological | Stirred fermentation tank verification, culture tank verification, process development data |
Food | Pasteurization verification, sterilization tunnel verification, rotary freezer verification, blast freezer verification, cold room verification |
Transport | Cold chain process tracking, cabin temperature detection |
Medical | High temperature sterilization testing, steam sterilization testing, medical supplies testing |
Technical Parameters
Temperature Verifier | Temperature and humidity tester | Pressure Verifier | ||
Model | TDLA | TDLC(Flexible probe) | MDLA | PDLA |
Operating temperature | -50 ~ 150 ℃(expandable to 400℃ with heat insulation box) | -50 ~ 150℃ (expandable to 400℃ with heat insulation box) | -40~125℃ | -40~140℃ |
Range | -50~150℃ | -50~150℃ | 0~100%RH | 0~100psi |
Resolution | 0.01℃ | 0.01℃ | 0.01℃ | 0.008psi; |
0.01%RH | 0.55mBar | |||
Accuracy | ±0.1℃ | ±0.1℃ | ±0.2℃ | ±0.25% Full scale |
±2%RH | ||||
Sensor | Pt100 Class 1/3 DIN | Pt100 Class 1/3 DIN | Temperature: Pt100 | Strain |
Humidity: capacitive | ||||
Capacity (recording points) | 64,000 | 64,000 | 8,000 | 8,000 |
Record frequency | 1 second to 18 hours | 1 second to 18 hours | 1 second to 18 hours | 1 second to 18 hours |
Battery Life | 36 months | 36 months | 12 months | 12 months |