Acrel AIM-M10 Nursing station Insulation Monitor for Medical IT system with CE
USD $1243.9 - $1244 /Piece
Min.Order:1 Piece
Jiangsu Acrel Electrical Manufacturing CO.,LTD
Technical consulting: Tina Li
Tel/Whatsapp:+86 15161626103
Email:sales02@email.acrel.cn
Data download
Product Introduction
AIM-M10 medical intelligent insulation monitoring instrument adopts advanced microcontroller technology, which has high integration, compact size, convenient installation and integrates intelligence, digitalization and networking in one. It is ideal selection for insulation monitoring of isolation power system in Class 2 medical locations such as operating room and intensive care unit.
AIM-M10 is usually combined with AITR series medical isolation transformer, AKH-0.66P26 current transformer and AID10/150 centralized alarm display device.
Functional DescriptionØ Functions of real-time monitoring and fault alarming of the ground insulation resistance, transformer load current and transformer winding temperature of the monitored IT system;
Ø Real-time monitor the temperature sensor disconnection fault and the functional grounding line disconnection fault of the monitored system, and give the alarm indication when the fault occurs;
Ø Relay alarm output, LED alarm indication and other faults indication functions;
Ø Adopt the advanced field bus communication technology and communicate the external alarm with the displayer to monitor operation condition of IT system;
DC24V power output function to provide the power supply for the external alarm and displayer.
Technical parameterAUX Power | Voltage | AC220V±10% | Temperature monitoring | Thermal resistor | 2 Pt100 |
Power consumption | ≤5VA | Measuring range | -50…+200℃ | ||
Insulation monitoring | Resistance measuring range | 10…999kΩ | Alarm value range | 0…+200℃ | |
Response value | 50…999kΩ | Alarm output | Output mode | 1 Relays | |
Relative uncertainty | ±10% | Contact rating | AC 250V/3A DC 30V/3A | ||
Response time | ≤3s | Environment | Operating temperature | -10…+55℃ | |
Permissible system leakage capacitance Ce | ≤5uF | Transport temperature | -25…+70℃ | ||
Measuring voltage Um | ≤12V | Storage temperature | -25…+70℃ | ||
Measuring current Im | ≤50uA | Relative humidity | 5%-95%,No condensation | ||
Impedance Zi | ≥200kΩ | Altitude | ≤2500m | ||
Internal DC resistance Ri | ≥240kΩ | IP degree | IP30 | ||
Permissible extraneous DC voltage Ufg | ≤DC280V | Rated impulse voltage / pollution degree | 4KV/Ⅲ | ||
Load current monitoring | Measuring Value | 2.1…50A | EMC/EMR | IEC 61326-2-4 | |
Alarm Value | 5…50A | Communication protocol | RS485,Modbus-RTU | ||
Measuring accuracy | ±5% |
Dimension Drawings
Front view Side view
Fig1 Dimension Drawings(unit: mm)
Wiring and Installation
Fig2 Terminal Diagram
1,2(FE,KE): Functional earthing
3,4(A,B): RS485 communication port
5,6(+24V,0V): DC power supply output
7,8(J1,J2): Relay output
11,12(L1,L2): IT system connector
15,16(I0,I1): Load current connector
17,18(T0,T1): Temperature sensor connector
Typical connection diagram
Certificate
Customers
Company Profile
Acrel Co.,Ltd.[Stock Code:300286.SZ] is a professional company in china who provides various systematic solutions of power monitoring,energy management and electrical safety for smart grid users.
Since June 2003,we have always been focused on the research,production and sales of intelligent electric meters.We concentrate on providing products and services to help customers improve electricity efficiency and safety.So far we have had the following comprehensive solutions for different applications:35(10)/0.4KV substation antomation system,power quality management system for medical location,precise distribution management system,smart lighting control system,remote orepayment metering system and motor control center.
Jiangsu Acrel Electrical Manufacturing Co.,LTD,our production base in Jiangyin City,Wuxi,provides guarantee for the industrialization and large-scare implementation of products.The annual production capacity of various meters is 2 million units,and more than 1.3 million units are sold in 2018.
Since the listing in 2012,our company has maintained steady growth.The company will adhere to the concept of“innovation,efficiency,unity,integrity”,adhere to steady operation and continuous innovation,and provide more high-quality products,services and silutions for power system users and micro-grid.
Contact