Blood And Infusion medical Warmer with CE
Negotiable /Piece
Min.Order:1 Piece
TECHNICAL DATA |
Name | infusion blood warmer |
Power supply | 100 - 240VAC/50-60 Hz |
Power consumption | 1.5A/115 VAC 1.0A/230VAC |
Fuse | 2×T2.5A/250V |
Type of protection against electric shock | Class I |
Degree of protection against electric shock | BF Applied part; Defibrillation-protected |
Degree of protection against ingress of liquids | IPX2 |
Temperature setting | 33°C ~ 41°C (91~106℉) ( 0.1°C increments) (1℉) |
Accuracy | ±1.0°C (±2.0℉) |
Overheat switch off | 42 ±1°C/43±2°C (108±2℉/110±4℉) |
Low temperature alarm | 32±1°C (89±2℉) |
Warming up time | Less than 2 minutes (23±2°C ~36°C) (73±4℉~97℉) |
Operating mode | Continuous |
MDD Classification | IIb according to rule 9 |
Dimensions: (L×W×H) | 184×148×110mm |
Warming profile | WP-143505, length 1.4 m, for infusion tube Æ3.5-5mm(O.D.) |
(Refer to the label on device) | WP-146505, length 1.4 m, for infusion tube Æ6-7mm(O.D.) |
Heating power | 60W |
Weight | 1.2kg |
Appearance and Control Panel |
Fig.1:General illustration |
Fig.2:Illustration of LED screen |
Fig.3:Illustation of the control panel |
The device can be fixed to the infusion stand or other support by using the clamp at the back.
Only use infusion stands or supports which be capable of supporting a minimum of ten pounds (4.5 kg) and with a minimum 60cm diameter wheelbase.
Make sure that they are sufficiently stable to against tipping over.
Warning:
After the device be fixed stable, it cannot be pull or application of force any way to avoid the infusion stands or supports tipping over. Failure to do so may result in damage to the device, the device may drop down with the infusion line and prevent or interrupt the infusion flow.
INSTALLATION AND OPERATION PROCEDURE |
Installation:
Check whether the device is damaged. Please do not use it if any damage has been found.
Fix the device firmly to the infusion stand at the position of 20-30cm below the infusion/blood bag, using the clamp at the back, then screw the fastening screw. The device is almost equal to the drip bucket.
Only use infusion stands or poles that are sufficiently stable.
Before connecting to mains power supply, check the voltage specified on the device label.
Connect power cable to power supply. The display lights on and beeps. The device is inSTANDBY mode.
Warning:
Do not locate the device at the position where it is difficult to be disconnected from
the mains power.
Fig.4:Illustration of the fastening clamp |
Operation procedure
Glide the roll clamp next to the drip bucket or at the end of the infusion tubing.
Empty the air in tubing to make infusion channel ready.
Take the primed infusion line at 2-3cm to the end of patient side to insert in the tail (small end) of the warming profile. Afterwards move your thumb over the infusion line and press theline gently into the groove towards the head (big end) of the warming profile, until the infusionline is completely wrapped in the warming profile.
Place the ends of the warming profile into the holder to prevent the infusion tube contacting with the ground.
Press the key or the key to select the desired temperature.
Adjust the appropriate infusion rate. As soon as infusion is started, press the key to start warming .
Resetting of temperature can only be performed in Standby mode.
If the infusion needs to be stopped for more than 3 minutes, press the key to stop warming. By press the key again to resume it.
Warning: The warming profile must not be placed directly on patient's skin to prevent scald by build-up heat.
Fig.5:Illustration of the holder of warming profile |
Shutting down the device
Press the key to enter STANDBY mode.
Release all pressure in the infusion system by switching off the pressure cuffs or infusionpumps.
Remove the consumable material from the warming profile and dispose it according to therelevant regulations.
Disconnect the device from the power supply. Clean and disinfect the equipment according to instructions.
Warning: If the over- temperature alarm is triggered or any suspected malfunction be perceived while in operation, the supply of liquid to the patient must be immediately stopped. Lock the connection tube to the patient, release all pressure in the infusion system by switching off the pressure cuffs or infusion pumps. Switch the device off or pull the mains plug out. The medium being used in the device must no longer be administered to the patient. Wait about 5 minutes until the warming profile has cooled down. Switch the device on again. If the fault occurs again, the device must be inspected by a trained service technician.