Anesthesia Machine/Anesthesia machine for ICU & operationg room
Negotiable /Piece
Min.Order:1 Piece
Gas requirement: Medical oxygen and nitrous oxide with a pressure ranging from O2: 0.32~0.6MPa ; NO2: 0.32 MPa to 0.6 MPa.
safety valve<12.5 kPa
Respiratory Rate 4~40bpm
Flow meter O2:0.1~10L/Min, N2O:0.1~10L/Min
oxygen concentration in the mixed gas N2O/O2 > 21%
Oxygen Flush: 25~75 L/min
Maximum error of the pressure testing device not exceed ± (4% of the full scale reading + 8% of the actual reading)
modes of ventilation A/C,IPPV, SIPPV, MANUAL
I/E Ratio: 1:1.5; 1:2.0;1:2.5; 1:3
Tidal Volume 0~1500ML
Controlled ventilation and assisted ventilation switching time 6s
Maximum safety pressure: ≤ 12.5 kPa
Pressure limit range: 0 ~ 6.0 kPa
Airway pressure alarm: Audible and visual and with yellow and red color indicating Lower: 0.2kPa~5.0kPa; Upper: 0.3~6.0 kPa
±0.2 kPa
Tidal volume alarm: Audible and visual and with yellow and red color indicating upper alarm: 50 to 2000ml,lower alarm: 0~1800ml
Power Supply Alarm "Ac/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s"