Pharmaceutical Packaging Water Vapor Permeability Testing Machine Infrared Method/ Electrolysis Method
Negotiable /Set
Min.Order:1 Set
Systester Instruments Co., Ltd
Blister Pack Oxygen Permeability Testing Machine Pharmaceutical Packaging
Weighing Method Water Vapor Permeability Testing Machine Decreasing And Increasing In One
Gurley Method Air Permeability Testing Machine Separator Membrane Test
Hydrogen High Pressure Permeate Test Gas Permeability Testing Machine Lab Test Apparatus
Plastic Bottle Cap Opening Force Torque Test Flexible Lab Testing Machine
Pharmaceutical Packaging Elongation/ Puncture Test Auto Tensile Testing Equipment UTM
Pharmaceutical Packaging Water Vapor Permeability Testing Machine Infrared Method/ Electrolysis Method
Products Details
Application:
Based on the electrolytic sensor method and available for the water vapor permeability test of flexible films, sheeting and packages. It can test 3 specimens in one operation with independent results.
Films:
Plastic films, composite films and many others.
Packages:
Plastic packages.
Pharmaceutical Packaging Water Vapor Permeability Testing Machine Infrared Method/ Electrolysis Method
Features:
— Electrolytic sensor method
— Own technology
— Film and package test
— 3 specimens test simultaneously and independently
— One-button test, test automatically in whole process
— Multiple test modes optional
— Data curves display
— Computer control
— Digital flow adjustment
— 3 Chambers with 3 temperature sensors
— Environment temperature and humidity sensor inside
— Chambers water bath technology
— Unit based on embedded system,24bit Δ-Σ AD
— Pressure, ppm water sensor and temperature overload protection
— Traceable reference film calibration
— Supports DSM system (DSM, lab data management system)
Pharmaceutical Packaging Water Vapor Permeability Testing Machine Infrared Method/ Electrolysis Method
Technical indexes:
Test range: | |
Film | 0.001-52 g/m2·24h (Standard) |
0.01-1100 g/m2·24h (Optional) | |
Resolution | 0.001 g/m2·24h |
Package | (optional) |
0.0001~0.5 g/pkg·d | |
Resolution | 0.00001g/pkg·d |
Specimen amount | 3 specimens, independent |
Temperature range | 5C – 95C |
Temperature accuracy | ±0.1C |
Humidity range | 0%RH,35%RH-90%RH,100%RH, standard condition 90%RH |
Humidity accuracy | ±1%RH |
Test area | 50 cm2 |
Specimen size | |
Film | size |
Thickness | less 5mm |
Package | |
Max size | D120mm * 400mm |
Carrier gas | 99.999% Nitrogen |
Carrier flow | 0~200 ml/min |
Gas interface | 1/8inch |
Products Catalog
what we can provide
Packages & Logistics
Export Areas
Contact