Plastic Composite Films Oxygen Permeability Values Gas Transmission Testing Machine
Negotiable /Set
Min.Order:1 Set
Systester Instruments Co., Ltd
Auto Peeling Force Tester Laminated Film Micro Peeling Testing Machine
BOPP/PV Lab Impact Testing Machine Plastic Film Pendulum Impact Tester
Gas Permeability Tester Lithium Battery Separation Membrane Gurley Air Permeability Tester
Gas Permeability Testing Machine Ultra-high Pressure Measurement Of Gas Transmittance
Heat Gradient Testing Machine Five Point Thermal Sealing Instrument Thermal Sealing Strength Tester
Thickness Tester PVC PE Membrane Thickness Gauge In Face Contact
Gas Transmission Testing Machine Plastic Packaging Oxygen Permeability Values Of Glass ASTM
Product Details
Application:
OTR-1901 is based on the coulometric sensor method and available for the OTR test of flexible films, sheeting and packages. It can test 3 specimens in one operation with independent results.
Films:
Plastic films, aluminum foil, aluminized films, plastic composite films, paper-plastic composite films, coextruded films, aluminum foil composite films and many others.
Packages:
Plastic, rubber, paper, paper-plastic composite, glass and metal packages, e.g. red wine bottles, blister, various plastic bottles, Coke bottles, peanut oil packages, Tetra Pak materials, vacuum bags, metal three-piece cans, plastic packages for cosmetic, soft tubes for toothpaste, jelly and yogurt cups.
Gas Transmission Testing Machine Plastic Packaging Oxygen Permeability Values Of Glass ASTM
Features:
— Coulometric sensor method
— Patent technology
— Film and package test
— 3 specimens test simultaneously and independently
— One-button test, test automatically in whole process
— Multiple test modes optional
— Data curves display
— Computer control
— Digital flow adjustment
— 3 Chambers temperature sensor embedded in
— Environment temperature and humidity sensor inside
— Chambers water bath technology
— Unit based on embedded system,24bit Δ-Σ AD
— Pressure, oxygen and temperature overload protection
— Traceable reference film calibration
— Supports DSM system (DSM, lab data management system)
Principles
Film:
The specimen is clamped between the upper and lower chamber, the upper chamber is purged by a oxygen stream, and the other side of specimen, lower chamber is purged by a stream of nitrogen, due to the effect of oxygen concentration gradient at two sides of the specimen, oxygen molecules will permeate through the specimen into the nitrogen side, penetrated oxygen molecules will be carried to the coulometric sensor which generates the proportional signal, so oxygen transmission rate and permeability coefficient will be obtained by calculating the sensor signal.
Package:
The package is mounted on the lower chamber, the outside of the package is air or a pure oxygen stream, the inside of the package is purged by a nitrogen stream, oxygen will permeate through the wall into the inside of the package, penetrated oxygen molecules will be carried to the coulometric sensor which generates the proportional signal, so oxygen transmission rate will be obtained by calculating the sensor signal.
Gas Transmission Testing Machine Plastic Packaging Oxygen Permeability Values Of Glass ASTchine
Technical indexes:
Test range: | |
Film | 0.01-6500 cm3/m2·d (Standard) |
0.07-6,5000 cm3/m2·d (Optional) | |
Resolution | 0.001 cm3/m2·d |
Package | |
0.0001~62 cm3/pkg·d | |
Resolution | 0.00001 cm3/m2·d |
Specimen amount | 3 specimens, independent |
Temperature range | 5C – 95C |
Temperature accuracy | ±0.1C |
Humidity range | 0%RH,35%RH-90%RH,100%RH |
Humidity accuracy | ±1%RH |
Test area | 50cm2 |
Specimen size | |
Film | size |
Thickness | less 5mm |
Package | |
Max size | D120mm * 400mm |
Carrier gas | 99.999% Nitrogen |
Carrier flow | 0~20 ml/min |
Gas interface | 1/8 inch |
Power supply | AC 220V 50Hz |
Dimension | 730mm * 590mm * 350mm |
Net weight | 57kg |