Features of Fetal Doppler Model: Sonoline C fetal doppler
Product Specification of Fetal Doppler
Safety: Complies with: IEC 60601-1: 1988+A1: 1991+A2: 1995 Classification: 1. Anti-electroshock Type: Internally powered equipment. 2. Anti-electroshock Degree: Type B applied part 3. Harmful Liquid Proof Degree: 4. Main unit: Ordinary equipment (sealed equipment without liquid proof). 5. Probe: Prevent from water splashing, degree of protection: IPX4. 6. Degree of Safety in Presence of Flammable Gases: Equipment not suitable for use in 7. Presence of flammable gases. 8. Working System: Continuous running equipment. 9. EMC: Group I Class B. 10. Suitable Using Range: Suitable for use after the 12th week of pregnancy.
Physical Characteristic 1. Size: 135mm (Length) × 92mm (Width) × 29 (Height) mm 2. Weight: About 245g (including batteries)
Environment Working: 1. Temperature: +5º C~+40º C 2. Humidity: ≤80% 3. Atmospheric Pressure: 70 kPa~l06kPa Transport and
FHR Performance 1. FHR Measuring Range: 50~240BPM (BPM: Beat per minute) 2. Resolution: 1BPM 3. Accuracy: ± 2BPM 4. Power consumption: <0.8W 5. Auto Shut-OFF: After 1 minute no signal, power off automatically. 6. Battery Type Recommended: Two pieces of 1.5 V DC battery (SIZE AA LR6).