transport incubator
USD $5000 - $5000 /Pair
Min.Order:1 Pair
Brief introduction:
1. Two temperature control modes: Air mode and Baby mode controlled by micro-computer;
2. Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance;
3. Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately;
4. Failure alarm indication;
5. Can be equipped with >37℃ temperature set function;
6. Double wall Acrylic glass hood with side door so that the bassinet can be pulled out from it;
7. Natural air flow humidity;
8. The height of whole unit can be adjusted;
9. Oxygen supply system;
10. Observation lamp
Standard configuration:
Main body (including the Acrylic glass hood, temperature controller, bassinet, internal battery, and observation lamp), oxygen cylinders, oxygen supply system, skin temperature sensor, I.V. Pole, mattress, adjustable stand)
Optional configuration: >37℃ override temperature mode
Specification
n AC Power supply: AC220V-230V, 50Hz
n DC Power supply: DC12V/10A or DC24V/6A
n Input power: 400VA
n Mode of control: Air mode and baby mode controlled by micro-computer
n Control range of air temperature: 25℃-37℃(Override mode 37-38℃)
n Control range of baby temperature: 34℃-37℃(Override mode 37-37.5℃)
n Accuracy of skin temperature sensor: ≤0.3℃
n Variability of temperature: ≤1.0℃
n Uniformity of temperature: ≤1.5℃
n Noise inside hood: ≤60dB(A)
n Failure alarm: over-temperature alarm, temperature deviation alarm, sensor alarm, Motor Fan Failed Alarm, Power failure Alarm and so on.
n Running time of internal battery: 90min (1 internal battery)
Requirements of environment
Environment temperature: 10 ℃~30℃
(Set temperature should be higher 3℃ than the environment temperature)
Wind speed of environment: ≤1.0m/s
Package
Package: each unit is packed as two cartons
Size (Main body): 109mm x 65.5mm x 76mm
(Cabinet): 138mm x 61mm x 40mm
Gross Weight (Main body): 80Kg
(Cabinet): 53Kg