LGSY208 Pressure Calibrator
Negotiable /Piece
Min.Order:1 Piece
Xian Wellwork Industrial Co., Ltd.
LGSY208 Digital Pressure Calibrator is a microprocessor controlled, precision instrument powered by an internal Li-ion battery or external 24V specified AC adapter, it is also has DC24V (80mV, accuracy 1%FS). It is portable and capable of accurate pressure instruments and sourcing when calibrating transducers, pressure switches, etc. The data storage is 6 months. The FSTN crystal screen is with deep blue color back vivid. It comprehends wide common-use, and provides many attached functions, such as leaking pressure checked, pressure button’s start, data recorded, convenient orders and pressure data appearances etc. It is closed to the modern international electronic instruments. LGSY 208 matches LGSY 205 pressure checked system to provide a set of calibration solution.
Function:
Calibration of pressure, electric and pressure digital without any hardware
Six high DC digital pressure, gauge inside.
Electric components are reliable highly because of the strict choice.
Less affection by temperature and humidity
Small but easily operate
Charge automatically
Special associated design.
Optional software for data processing
A charge may use for three months (with 24v output is shorter)
Pressure record includes pressure, environmental temperature and time parameters etc.
9 pressure unit can be selected indicators With pa, KPa, MPa or psi, KGF, bar and mbar, mmHG, mmH2O
- 100kpa ~2KPa ~ 250kpa within the media is only the uncorrosive gas. The range is 0.4MPa~250MPa and the pressure is from stainless steel 316l compatible with the fluid or air;
Regular pressure range chose
Mark:
Pay attention to differential and absolutely calibrator when ordered.
Pressure Measured Accuracy:±0.05%F.S;±0.025%F.S.
Pressure Unit:mmH2O, mmHg , psi, kPa, Mpa, pa, mbar and bar. Different range matches different units.
Pressure over-loaded alarm: Over 120%, rhythmic sounds buzzed.
Electric Measured Characteristics:
Electric Measured:±30.0000mA;Accuracy:±(0.02%RD+0.003%FS);
Electric Measured:±30.0000V; Accuracy:±(0.02%RD+0.003%FS);
Temp Measured: It tests for the pressure transmitter’s temperature.
Switch Measured: Confirm it is on or off. If with checked pressure, the voltage is 1-12V.
Output: DC24V±0.5V; Max output:80mA;
Environmental:
Temp: (0~50)°C(Assure the exact working temp scale)
B. Humidity:<95%;
C. atmospheric pressure:(86~106)KPa;
Calibration Temp:20°C±2°C;
Storage Temp: -20°C~70°C;
Overall dimension:φ120mm×40mm;
Weight:0.9kg
Electrical conn;
Measured Conn:Standard Jack 2mm
Output Conn:Standard Jack 2mm
Charge Conn Standardφ4 charger
RS232 Conn : Standard DB9 charger
Pressure connection:M20×1.5