Electric medical test equipment hematology cell counter
USD $4,000 - $5,000 /Set
Min.Order:1 Set
Guangzhou Joyful Technology Co., Ltd.
Electric medical test equipment hematology cell counter
Changes in human physiology and pathology often lead to differentiation in blood components, such as infectious inflammation of whole body tissues, which could cause blood leukocytes count changes. So to keep abreast of changes in blood components, hematology analyzer can provide diagnosis results as doctor’s important basis to treatment. The blood cell count is a count of unit volume, mainly red blood cells, white blood cells and platelets number, it has become one of the hospital at all levels for routine blood tests.
Specification
Model |
TE-3000 |
Parameters |
23Parameters 20+3histogram |
Principles |
Impedance |
Sample volume |
20μl venous blood or capillary |
Sample processing |
Open tube |
Aperture Diameter |
WBC 80 μm RBC/PLT 80μm |
Dilution Rations |
Whole blood 1-40000 Capillary Blood 1-40000 Whole blood 1-500 Capillary Blood 1-250 |
Data storage |
20000 test results with histogram |
Throughput |
60 samples/hour |
Display |
8 inch colour LED |
Carryover |
WBC,RBC,HGB≤0.5%,PLT≤1% |
Aperture |
Double apertures for WBC and RBC/PLT |
Patented aperture technology |
Automatic block-clearing of high pressure and lower rate of block |
Printout |
built-in the thermal printer(optional external printer) input/output one RS232,1keyboard,1USB,1printer port |
Operating environment |
Temperature 15°C-30°C,Humidity 30%-85% |
Power requirement |
100-240V~50/60Hz |
Dimension |
440(W)*354(D)*430(H) |
Weight |
17KG |
Feature
Technical parameter
1.How It Works: Electrical impedance method, cyanide methemoglobin method and non-cyanide SFT method
2.Test Parameters: WBC three categories, 22 parameters, 23 parameters
(including color histograms of WBC, RBC, PLT)
3.Specimen Amount: Whole blood: 9.6μL; peripheral blood(direct detection): 9.6μL; peripheral blood
(pre-dilution mode): 20μL
4.Test Speed:Single-channel: 30 samples/h, dual-channel: 60 samples/h. May be continuous 24 hours a day, with automatic sleep and wake functions
5.Results Storage: The device saves 1.2 million copies of the results with all three histograms automatically
6.Quality Control Methods:L-J graph, X, SD, CV%
7.Automatic detection reagent margin, closed reagent tube
8.Calibration: Manual calibration and automatic calibration
9.Reference Value Setting: 8 settings groups with adult male, adult female, child, newborn, general all purpose and 3 groups of custom settings
10.Information: The device supports a variety of multiple patient information fast Chinese input, including 12 kinds of information such as name, department and censorship.
11.Operation: Chinese Windows color operating system, graphical buttons, supporting mouse operation
12.Interface: A parallel printer interface, a VGA display interface, two RS-232 interfaces, two USB interfaces and two PS / 2 interfaces
13.Report Printing: Built-in high thermal printing, optional six reporting formats in English, and an optional external printer
14.Working Environment: Working temperature of 15-35°C, and can be used in high altitude environment
15.Power supply: AC 220V 50HZ
Warranty
There are 1 years ftee maintenance and more than one year life time for machine .
We guarantee quick and efficient service and maintenance of our device security solutions around the world. We ensure that we will be there whenever or wherever you need us, 24 hours a day, 365 days a year. Like all Onity solutions, our products are backed by a commitment to provide superior global service with local support. we can help you work and reach your goals.