BKJ6100 Auto Hematology Analyzer
Name and the using range
XFA6100 Automatic Hematology Analyzer is mainly used to detect the blood series parameters of red blood cells and hemoglobin, platelet and leukocyte. It is a vitro diagnostic equipment which is applied to quantitative analysis of blood cells of clinical and laboratory.
Principle of cell counts
The instrument measured blood can be diluted to the exact proportion by diluents firstly, which is used to measure the parameters of WBC,RBC,PLT and HGB. WBC, RBC counting pool diluted samples, then pass their own porous counting respectively under negative pressure, the amount of fluid that flowing through the small hole is the specimen volume of the instrument for analysis.
The instrument measured the size distribution and number of WBC, RBC and PLT by using impedance method. Electrodes inserted in the holes on both sides of the liquid, it sets up a field environment on both sides of the surrounding count micro porous, according to the blood-cell non-conductive nature, when the cells passed the count micro porous, it will cause impedance changes, the size of the impedance changes and proportional to the cell volume.
Specifications:
Test Items | WBC, LY#, MO#, GR#, LY%, MO%, GR%, HGB, RBC, HCT, MCV, MCH, MCHC, RDW—CV, RDW—SD, PLT, PCT, MPV, PDW, WBC Histogram, RBC Histogram, PLT Histogram | ||
Background Count | WBC: ≤0.3×109/L RBC: ≤0.03×1012/L HGB: ≤1g/L PLT: ≤10×109/L | ||
Linearity | Test Item | Test Range | Allow Difference of Linearity |
| WBC | 0∼6.0×109/L | ±0.3×109/L |
6.0∼99.9×109/L | ±5% | ||
RBC | 0∼0.99×1012/L | ±0.05×1012/ L | |
1.0∼9.99×1012/L | ±5% | ||
HGB | 0∼99 g/L | ±2g/L | |
100∼300 g/L | ±2% | ||
PLT | 0∼99×109/L | ±10×109/L | |
100∼999×109/L | ±10% | ||
Repeatability | Test Item | CV(%) | Teat Range |
WBC | ≤2.0 | 7.0~15.0×109/L | |
RBC | ≤0.5 | 3.5~6.0×1012/L | |
HGB | ≤1.5 | 110~180g/L | |
PLT | ≤4.0 | 150~500×109/L | |
MCV | ≤0.5 | 80~110fL | |
Sample Volume | Twig blood: 20μl | ||
Dilution Ratio | WBC/HGB: 1:500 RBC/PLT: 1:50000 | ||
Interfaces | RS232 joint, USB, PS2, Parallel port | ||
Test Speed | 60 samples/hours | ||
Continuous Work Time | >8h |