Elisa Analyzer/ Immunoassay System (skype: fangfeimengxiang876)
USD $9,000 - $10,000 /Unit
Min.Order:1 Unit
Zhangqiu Meihua International Trading Co., Ltd.
Elisa Analyzer/ Immunoassay System
BIOBASE Elisa Processor is fully automated two-plate ELISA processing system capable of adding sample,incubation,washing and data processing.
It is complete solution for your IVD ELISA assays. The system combines a high precision pipetting unit with stable incubators and efficiency washers and photometer unit.
BIOBASE1000 Modularity:
1 Robotic hand.
1 Pipetting probe, 1ul stepping (1000ul volumes).
80 Sample rack, 92 Dilution rack.
2 units 96 well microplate(pipetting,incubator,reading).
1 unit independent washer&reader( two 8 nozzle manifolds).
Testing Range:
1. TORCH SERIES 4. THYROID AUTOIMMUNITY SERIES
Toxo, RV, CMV, HSV etc. T3, T4, FT3, TSH etc
2. HEPATITIS SERIES 5. ENDOCRINE SERIES
HAV, HBV, HCV, HDV, HEV etc. E2, E3, LH, FSH, HCG, Pro, PRL etc.
3. VENEREAL SERIES 6. TUMOR MARKER SERIES
HIV, TP, MP, CT etc. CEA, AFP, PSA F-PSA, CAL9-9, CAL 25, CA15-3, CA242.
Modularity Composition:
2 microplate ,80 Sample rack, 92 Dilution rack.
All-in-one microplate reader and washer
8 special reagent racks, 1 general reagent racks.
Robotic precision probes and automated washing system.
Washer:
Two 8 nozzle manifolds
Automatic liquid level sensor detection
Separate aspiration and dispense probes prevent cross contamination
Reader:
8 independent photometric channels
Wavelength range 400-700nm
2 precision filters (450,630nm), 2 more available on request.
User-friendly Software:
LIS system available (laboratory information system)
Large data storage capacity
Professional service software monitor process on line status.
Technical Parameters:
Model | Biobase1000 | |||
Sample pipetting unit | Sample rack | 80 sample tubes(12 mm) | ||
Primary sample tubes | 12*75mm | |||
Pipetting | 8ul~1000ul, (10ul) CV<5% | |||
Dispensing system | 1 aspirating-dispensing probe with X-Y-Z movement;dilution system with a syringe of 1000ul, continuous flow washing station. | |||
Reagent pipetting unit | Reagent racks | 8 special reagent racks,1 general reagent racks | ||
Removable trays | modular racks to hold calibrators, controls and reagents for 8 test on line | |||
Pipetting | 8ul~1000ul,1ul stepping, | |||
Pipetting precision | (100ul )CV≤0.2~1% | |||
Time to dispense | 4 minutes(typical)^100ul of 96 reagents to plate from reagent bottle | |||
Incubation unit | Incubators | 2 independent units | ||
Temperature range | RT to 45 °C | |||
Temperature accuracy | ±0.2°C | |||
Adjustment step size | ±0.1°C | |||
Washing unit | Programmable | Yes(options & parameters) | ||
Wash containers | 2 wash buffers at 2.0L, with level-sensing | |||
Washing system | two 8 nozzle manifolds, one to dispense and the other one to aspirate; | |||
Washing residual | <2µl | |||
Reading unit | Reading system | 8 independent photometric channels with mono and dichromatic reading capability | ||
Photometric range | 0.000 ~3.000 OD | |||
Spectral range | 400-700nm | |||
Optical filters | 2 precision filters (450,630nm); 2 more on request (4 filters max, 400-700nm) | |||
Software unit | Software system | Windows XP or Windows 7 or above, LIS system available | ||
Function application | work list set up, patient reports and result filing; calibration curve storage for each carried out test | |||
Data reduction | cut-off (qualitative) | |||
Physical Specification | Dimensions | 790x610x650mm (W*D*H) | Weight | 80kg |
Power Supply | AC110/220V+ 10%,50/60Hz | Consumption | 500W |