High Performance Biochemical Multipurpose Analyzer- 300 test with CE,FDA certifiate (skype: fangfeimengxiang876)
USD $6,200 - $6,500 /Set
Min.Order:1 Set
Zhangqiu Meihua International Trading Co., Ltd.
High Performance Biochemical Multipurpose Analyzer- 300 test with CE,FDA certifiate
Application:
The biochemistry analyzers are multi-functional,can run many test items.
Following are test details:
1. Liver Functions Test: ALT, AST, TP, ALB, TBIL, DBIL, ALP, CHE, TBA
2. Blood Lipid Items: CHOL, TG, HDL-C, LDL-C, and Apo A1, Apo B, AMY, LIP, LP (a)
3. Myocardium Zymogram Items: LDH, CK, CK-Mb,ACE
4. Blood Gas Electrolyte Items: K+, Na+, Cl-, Ca++, Mg++, P
5. Kidney Function Items: UREA, UA, CREA, MALB
6. Glycometabolism Items, Pancreas Items, Pepsinogen Items, etc.
Overall performance:
Model: CRYSTAL
Type: discrete, fully automatic
Throughput: 300T/H with single reagent
Analysis Method: end-point, fixed-time (two-point),kinetics,turbidimetry
Programming Items: programmable for any test items
Certification: CE, FDA, ISO9001, ISO14001, ISO13485
1.Sample unit:
Sample Settings: 60 sample positions, including calibration, quality control and emergency, primary tubes and serum is available
Sample Volume: 2ul ~50ul (0.1ul stepping)
Sample Probe: Having liquid level induction function, ensuring the accuracy of adding sample
Probe cleaning: Automatic washing both interior and exterior
Analysis Items: analyzing maximally 45 items at the same time
2.Reagent unit:
Reagent Position: 45 low-temperature refrigerated reagent position
Reagent Volume: R1:25~300ul, R2:10~150ul (1ul stepping), supporting reagent from different countries
Temperature Control: continuous 2C~8Ccooling within 24 hours
Reagent Probe: Having liquid level induction function, ensuring the accuracy of adding reagent. Inventory checking
Probe cleaning: Automatic washing both interior and exterior
3.Reaction unit:
Reaction Time: 10 minutes
Cuvettes: 120 independent reaction cuvettes (optical path length 6mm)
Stirrer: after injecting reagent and sample, blending immediately
Cleaning: automatic water cleaning and drying system
Temperature Control: 37C±0.1C
System Adjustment: system can record the blank value and get more accurate result by deducting blank base value of each cuvette
4.Optical System:
Light Source: 12V/20W halogen-tungsten lamp, 2000 hours
Receiver: 8 imported photo receivers
Wavelength Range: 340,405,450,510,546,578,630,700(three of them can be defined by users), adopting reversed optics
Testing Wavelength: single and double wavelength test
5.Calibration and QC:
Calibration: one-point, multi-point
Linear Range: 0.000~2.500Abs
Quality Control: Inserting quality control at random.6 quality controls can be reached with one item
6.Software:
Software: Windows 2000 or Windows XP or above, sample and easy to operate
Emergency Function: emergency sample can be placed in the sample tray at any time
Data Processing: Having reaction curve detection and data display function in the whole process
Report Printing: report format can be modified by user
7.Others:
Working Temperature: 15C~30C
Humidity: 35%~80%
Power Supply: AC220V+22V 50Hz+1Hz
Dimension:1020mm×830mm×1200mm (W*D*H)
Weight: 130kg