Ceramic Thermocouple
USD $1 - $10 /Piece
Min.Order:1 Piece
Shaoxing Shangyu Shenzhou Instrument Co., Ltd.
Thermocouple are used in industry for measuring the temperature of gases, liquids or solid surfaces etc., Its usually assembled with indicators, recorders or computers.
A thermocouple consists of two dissimilar metals joined together at one end. When the junction of the two metals be cooled or heated and a voltage is produced that can be correlated back to the temperature,
Shangyu Shenzhou Instrument Co., Ltd providing Thermocouples are manufactured in different combinations of metals. The calibrations are K, N, E.J, T, S, R, B. Each calibration has a different temperature range.
TYPE AND SHEATH TABLE:
Product Name | Code | Type | Sheath material | Outer dia.(mm) | Temperature |
Ni Cr - Ni Si | KK | K | SS304 | 0.5-1.0 | 400 |
SS321 | 1.5-3.2 | 600 | |||
SS316 | 4.0-8.0 | 800 | |||
| 0.5-1.0 | 500 | |||
SS310 | 1.5-3.2 | 800 | |||
INCL600 | 4.0-6.4 | 900 | |||
| 8.0-12.7 | 1000 | |||
Ni Cr Si - Ni Si | NK | N | SS304 | 0.5-1.0 | 400 |
SS321 | 1.5-3.2 | 600 | |||
SS316 | 4.0-8.0 | 800 | |||
| 0.5-1.0 | 500 | |||
SS310 | 1.5-3.2 | 800 | |||
INCL600 | 4.0-6.4 | 900 | |||
| 8.0-12.7 | 1000 | |||
Ni Cr -Konstantan | EK | E | SS304 | 0.5-1.0 | 400 |
SS321 | 1.5-3.2 | 500 | |||
SS316 | 4.0-8.0 | 800 | |||
Fe - Konstantan | JK | J | SS304 | 1.0 | 300 |
SS321 | 1.5-3.2 | 500 | |||
SS316 | 4.0-8.0 | 800 | |||
Cu - Konstantan | TK | T | SS304 | 1.0 | -200-100 |
SS321 | 1.5-3.2 | 100-200 | |||
SS316 | 4.0-8.0 | 100-350 |
ACCURACTY TABLE:
Name | Type | Class of Accuracy | |||||
I | II | III | |||||
Tolerance | Temperature | Tolerance | Temperature | Tolerance | Temperature | ||
Ni Cr - Ni Si | K | 1.5 or 0.4%t | -40~1000 | 2.5 or 0.75%t | -40~1000 | 2.5 or 0.75%t | -200~40 |
Ni Cr Si- Ni Si | N | -40~1000 | -40~1000 | -200~40 | |||
Ni Cr - Konstantan | E | -40~800 | -40~800 | -200~40 | |||
Fe - Konatantan | J | -40~750 | -40~750 | --- | |||
Cu - Konstantan | T | 1.5 or 0.4%t | -40~350 | 1 or 0.75%t | -40~350 | 1 or 0.75%t | -200~40 |
Pt Rh 10 - Pt | S | 1 or 1+(t-1100)x0.003 | 0~1600 | 1.5 or 0.25%t | 0~1600 | --- | --- |
Pt Rh 13 - Pt | R | 0~1600 | 0~1600 | ||||
Pt Rh 13 - Pt Rh 6 | B | --- | --- | 600~1700 | 4 or 1.5%t | 600~1700 |
Thermocouple is for measuring temperature, the tube with double or single crematic materials
We offer products and services to meet any need at any technology level.
Supply and services:
Thermocouple wire made of platinum rhodium 13-platinum rhodium 6, the wire in conformity with standard GB/T1598-2010
Technical Data
Thermocouple specification | Insulation resistance | Accuracy | |||||
419.527°C | 660.323°C | 961.78°C | |||||
130-1 R Φ15*600*400 | 3620 | 6270 | 9989 | ≥500MΩ | Ⅱ | ||
130-2 R Φ15*600*400 | 3623 | 6275 | 9995 | ≥500MΩ | Ⅱ | ||
130-3 R Φ15*600*400 | 3617 | 6266 | 9987 | ≥500MΩ | Ⅱ |