TH2827C,high frequency lcr meter,Equal to Agilent E4980AL Lcr Meter,digital lcr meter,1MHz
Negotiable /Piece
Min.Order:1 Piece
Xiamen Wellzion Electronics Co., Ltd.
TH2827B,high frequency lcr meter,Equal to HP E4980AL LCR Meter,500kHz Lcr meter
TH2827B,high precision lcr meter,Equal to Agilent E4980AL LCR Meter,500kHz Lcr
TH2829C,Automatic Component Analyzer,1MHz Component Analyzer
TH2829B,Automatic Component Analyzer,500kHz Component Analyzer
TH2829A,Automatic Component Analyzer,300kHz Component Analyzer
Display | 480*RGB*272 4.3-inch TFT LCD display | |
Frequency of test signal | ||
TH2827A | 20Hz ~ 300kHz | |
TH2827B | 20Hz ~ 500kHz | |
TH2827C | 20Hz ~ 1MHz | |
Minimum resolution | 10mHz, 4-digit frequency input | |
Accuracy | 0.01% | |
AC Level | ||
Voltage range of test signal | 10mV ~ 2Vrms | |
Minimum resolution of voltage | 100μV, 3-digit input | |
Accuracy | ALC ON | 10% x set voltage + 2mV |
ALC OFF | 6% x set voltage + 2mV | |
Current range of test signal | 100μA ~ 20mA | |
Minimum resolution of current | 1μA, 3-digit input | |
Accuracy | ALC ON | 10% x set current + 20μA |
ALC OFF | 6% x set voltage + 20μA | |
DC bias voltage source | ||
Voltage/Current range | 0V ~ ±5V / 0mA~ ±50mA | |
Resolution | 0.5mV / 5μA | |
Voltage accuracy | 1% x set voltage + 5mV | |
ISO ON | Be used for the bias test of inductance and transformer | |
AC Source impedance | ISO ON | 100Ω |
ISO OFF | 30Ω,50Ω,100Ω selectable | |
DCR source impedance | 30Ω,50Ω,100Ω selectable | |
Test parameters of LCR | |Z|, |Y|, C, L, X, B, R, G, D, Q, θ, DCR, Vdc-Idc | |
Parameter display of test page | One set of main/sub parameter, 10-point list sweep | |
Test parameters of transformer | DCR1(primary, 2-terminal), DCR2(secondary, 2-terminal), M (mutual inductance), N, 1/N, Phase, Lk(leakage inductance), C(primary, secondary capacitance), balance test | |
Basic accuracy | LCR test parameter | 0.05% |
N | 0.10% | |
Calibration | Warm-up time ≥ 30 seconds; Environment temperature: 23±5ºC; Signal voltage: 0.3Vrms-1Vrms; Zeroing: After OPEN or SHORT; Length of test cable: 0 m | |
Measurement time (≥10 kHz) | Fast: 13 ms / time Medium: 67 ms / time Slow:187 ms / time Plus the refresh time of display character | |
Display range of LCR parameter | | Z | , R ,X,DCR | 0.00001Ω ~ 99.9999MΩ |
|Y|,G,B | 0.00001μs ~ 99.9999s | |
C | 0.00001pF ~ 9.99999F | |
L | 0.00001μH ~ 99.9999kH | |
D | 0.00001 ~ 9.99999 | |
Q | 0.00001 ~ 99999.9 | |
θ(DEG) | -179.999º ~ 179.999º | |
θ(RAD) | -3.14159 ~ 3.14159 | |
Δ% | -999.999% ~ 999.999% | |
Equivalent circuit | Serial, Parallel | |
Range mode | Auto, Hold | |
Trigger mode | Internal, Manual, External, Bus | |
Average times | 1-255 | |
Calibration function | Open,short calibration with full frequency or dot frequency,Load | |
Math operation | Direct reading, ΔABS, Δ% | |
Delay time setup | 0-999, minimum resolution: 100us | |
Comparator | 10-bin sorting, BIN1-BIN9, NG,AUX | |
Bin counter | ||
PASS/FAIL on front panel,LED indication | ||
List sweep | 10 points list sweep; Frequency, AC voltage/current,internal/external bias voltage/ current can be swept; Each sweep point can be sorted separately; | |
Internal nonvolatile memory | 100 sets of LCRZ setting files 201 times test results | |
External USB memory | GIF files LCRZ setting files Test data can be stored via USB memory directly. | |
Interface | I/O interface | HANDLER on rear panel |
SCI | USB, RS232C | |
PCI | GPIB (optional) | |
NI | LAN | |
Memory interface | USB HOST (front panel) | |
Bias current source control interface DCI | External DC bias current source can be controlled by using DCI interface. The maximum bias current can reach 120A. | |
Option. Only choose either DCI or GPIB | ||
General Specifications | ||
Operating temperature and humidity | 0ºC ~ 40ºC, ≤ 90%RH | |
Power supply | Voltage | 99V-121V, 198V-242V AC |
Frequency | 47Hz-63Hz | |
Consumption | Max. 80 VA | |
Dimension(W*H*D) | 280 mm * 88 mm * 370 mm(with no sheath) 369 mm * 108 mm * 408 mm(with sheath) | |
Weight | Approx. 5 kg |
Model | Agilent E4980AL | TH2827 Series |
Frequency | 20 Hz to 300 kHz/500 kHz/1 MHz | 20 Hz to 300 kHz/500 kHz/1 MHz |
Test signal level | 0 to 2 Vrms/0 to 20 mArms | 10mV ~ 2Vrms/ 100μA ~ 20mA |
Range mode | Auto, Hold | Auto, Hold |
Trigger mode | Internal, Manual, External, Bus | Internal, Manual, External, Bus |
Programmable list sweep | 201 points | 10 points |
Remote control | GPIB, LAN, USB | GPIB(optional), LAN, USB |
Delay time setup | 0-999, minimum resolution: 100us | 0-999, minimum resolution: 100us |
Average times | 1 - 256 | 1-255 |
Comparator | 8-bin sorting;Bin counter | 10-bin sorting;Bin counter |
Equivalent circuit | Serial, Parallel | Serial, Parallel |
Basic accuracy | 0.1 % @ SHORT 0.05 % @ MED/LONG | 0.1%@ N 0.05%@ LCR test parameter |
Parameters | Cp-D/Q/G/Rp,Cs-D/Q/Rs,Lp-D/Q/G/Rp,Ls-D/Q/Rs,R-X,Z-Ød/Ør,G-B,Y-Ød/Ør,Lp-Rdc, Ls-Rdc | |Z|, |Y|, C, L, X, B, R, G, D, Q, θ, DCR, Vdc-Idc |
Storage devices | Internal/USB memory | Internal/USB memory |
Cabinet dimensions (mm) | 370 (W) x 105 (H) x 390 (D) mm | 280 mm * 88 mm * 370 mm |
Weight | 5.3 kg | Approx. 5 kg |