TECHNIQUES SPECIFICATIONS

Physical specifications

Display

 10.4'' LCD Display

Mode:

pneumatically powered and electrically controlled system

Trolley:

Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions

Technical Specifications:

Gas requirement:

Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpa

safety valve

<12.5  kPa

Respiratory Rate

1 ~ 99 bpm

gas concentration output

21% ~ 100%  adjustable

Maximum Minute ventilation

18L / min

modes of ventilation

A/C, SIPPV,IPPV,SIMV,PSIMV,CPAP,PCV,PEEP,SIGN,SPONT, MANUAL

Ventilator compliance

30 Ml/kPa

PEEP:

0 ~ 1.0 kPa

Ptr:

-10 ~ 10 hPa

IMV frequency:

1~20 bpm

I/E Ratio:

4:1~1:4, Has the inverse ratio ventilation

Tidal Volume

Setting Range:  50~1500ML   Display Range: 0~2000ml ;  resolution ratio: 10ml

Inspiratory Plateau:

0~1s

Maximum Safety  Pressure

≤ 6.0kPa

Sigh:

one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point

Pressure limit range:

1.0~ 6.0kPa

Airway pressure alarm: Audible and visual and with yellow and red color indicating

Lower: 0.1 kPa ~ 6.0 kPa;   Upper: 0 ~ 5.0  kPa

±0.2 kPa

Ventilation Volume alarm:

≥ 18L/ MIN

Power Supply Alarm

Ac/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s

The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.

 

Medical Anesthesia Machine,Anesthesia Equipment used in Hospital,Clinical

 

Hot Sale

Medical Anesthesia Machine,Anesthesia Equipment used in Hospital,Clinical 

Our Services

1. OEM/ODM service : Darppon offers OEM/ODM service for distributor to establish long term business relationship.

 

2. Warranty :Darppon  offers 18 months free after-sale service for customers, and be responsible   for life-time repair.

 

3. Technical support:Darppon offers professional engineers for customers on the technical support, and response within 12 hours

Contact Information

Medical Anesthesia Machine,Anesthesia Equipment used in Hospital,Clinical 

Packaging & Shipping

Medical Anesthesia Machine,Anesthesia Equipment used in Hospital,ClinicalMedical Anesthesia Machine,Anesthesia Equipment used in Hospital,Clinical 

 

Contact Supplier

Mr. Edan Hu manager Chat Now
Telephone
0086-27-87053356
Mobile
18602700936
Address
Room B2713, Optical Valley Time Square, East Lake High-Tech. Development Zone Wuhan,Hubei

You May Like

New Products

Recommended Products

View More View Less

Find Similar Products By Category

Health & Medical > Medical Devices > Emergency & Clinics Apparatus
Categories
Hi~