Specifications
Automatically detection of syringe types and settings.
Broad range in delivery rate settings.
8 Hours internal battery life
FEATURES
Automatically detection of syringe types and settings.
Broad range in delivery rate settings.
8 Hours internal battery life in high capacity.
Three levels of alarm volume: high, medium and low.
Three levels of occlusion pressure: high, medium and low.
KVO function After infusion is complete, the pump automatically switches to the smallest flow to prevent creating the thrombus
.
SPECIFICATIONS
Compatible Syringe: Any standard syringe with 10ml,20ml(30ml) and 50ml(60ml)
Accuracy +-2% (Any manufacturers of syringes, right after calibration)
Delivery Rate Range
50ml(60ml) 0.1ml/h~1800.0ml/h (0.1ml/h step)
20ml(30ml) 0.1ml/h~800.0ml/h (0.1ml/h step)
10ml 0.1ml/h~400.0ml/h (0.1ml/h step)
Time Mode Range
Volume 0.1~999.9ml (0.1ml step)
TIME 1~1999min (1min step)
Body Weight Mode Range
Volume 0.1~999.9ml(0.1ml step)
DRUG MAS0.1~999.9mg(0.1mg step)
DOSE 0.01~99.99ug/kg/min(0.01ug/kg/min step)
WEIGHT 0.1~300kg(0.1kg the least step)
Volume Limit Range 0~1999.9ml (0.1 ml step)
Total Volume Range 0.1~1999.9ml (0.1 ml step)
Rapid injection: 50ml(60ml) 1800.0ml/h, 20ml(30ml) 800.0ml/h, 10ml 400.0ml/h
KVO Rate 1ml/h
Alarms Near empty, Syringe empty, Volume complete(a volume limit has been set ), Occlusion, Syringe loose, Operation error, Low battery and other device failures.
Power Supply AC power supply: AC 100 to 240V±10%,50/60Hz
Internal Lithium Battery 11.1V, 2000mAh. Approximately 8 hours of continuous operation on a new and fully charged battery running at 5ml/h.
Power Consumption: ≤25VA
Safety level Class 1, Type CF
Dimensions 280mm(L)×160mm(W)×130mm(H)
Net Weight Approximately. 3.0kg
Operating Conditions Ambient temperature 5 to 40°C Relative humidity 20 to 90%
Storage Condition Ambient temperature -10 to 55°C Relative humidity 10 to 90%
Accessories AC power line, Operational manual