Syringe pump used for ward
USD $200 - $300 /Piece
Min.Order:1 Piece
1. Flow rate:
0.1ml/h-1200ml/h(0.1ml/h increments,1ml/h increments when rate is over 1000ml/h )
50ml syringe
0.1ml/h-600ml/h(0.1ml/h increments) 30ml syringe
0.1ml/h-399.9ml/h(0.1ml/h increments) 20ml syringe
2. Bolus rate: 1200ml/h (50ml syringe); 600ml/h (30ml syringe); 399ml/h (20ml syringe)
3. Occlusion pressure: 40N, 60N (±10N)
4. Volumetric accuracy: ≤2% (drive accuracy ≤1%)
5. Total volume indication: 0.1-9999ml
6. Preset limit volume: 0.1-9999ml
7. Alarm: Near empty, Empty, Syringe dislocated, Syringe plunger disengaged, Malfunction, Forget to operate after power on, Occlusion, Low battery, Battery volume exhausted, AC power failure, preset limited volume is finished
8. Operating condition: Temperature:-5 to 40; Relative humidity: 20 to 90%
9. Transport and storage condition: Temperature:-40 to 55
10. External dimensions: 307(L) ×135(W) ×128(H) mm
11. Net weight: 2.1kg (include internal battery and I.V. Clamp)
12. Syringe: Syringe size: 20ml/30ml/50ml; syringe type: 13 kinds of disposable brands can be used
13. Range of body weight mode (For JAZ -6T):
Dose rate: 0.01-99.99 u g/kg/min (0.1 u g/kg/min increments)
0.01-99.99mg/kg/h (0.01 mg/kg/h increments)
Body weight: 0.1-300kg (0.1kg increments)
Drug volume: 0.1-999.9mg (0.1mg increments)
Solution volume: 0.1-999.9ml (0.1ml increments)
Other functions:
Body weight mode(for WZ-506CT only),
occlusion pressure selected,
anti bolus,
syringe brand selected,
automatically recognize syringe size,
preset limited volume,
total volume indicate,
KVO,
fast-injection control,
lock the fast key when infusing,
lock the flow rate keys when infusing,
audio and video alarms,
AC and DC indicate,
RS232 port,
warn of wrong operation,
500 log.