Hipot and Insulation Resistance Tester for PV Module
Negotiable /Unit
Min.Order:1 Unit
Nanjing Allwin Instrument Science And Technology Co., Ltd.
The CS9922G CS9922G-1, AC/DC Withstanding Voltage/Insulation Resistance Tester, is the test instrument that can handle the insuration test with high voltage and high resolution required for the evaluation of the PV module, Cable, Connector, and Junction Box. The CS9922G/9922G-1 are equipped with functions of the AC/DC withstanding voltage testing and the insulation resistance testing accompanied with the features and performance of IEC61730, and both of the two extends additional features and specifications exclusively applied to the PV module testing. Furthermore, the CS9922G, CS9922G-1 improve the current measurement accuracy of the DC withstanding voltage testing from the original 0.100mA to the current 0.010uA.
Model | CS9922G | CS9922G-1 | ||||||
ACW | Output voltage | range | 0.050kV~5.000kV | |||||
accuracy | ±(1.5%read value +5V) | |||||||
resolution | 1V | |||||||
Max output power | 100VA(5.000kV/20mA) | 500VA(5.000kV/100mA) | ||||||
Max test current | 20mA | 100mA | ||||||
Low-current | 0~20mA,0=not judgment | 0~100mA,0= not judgment | ||||||
Range | 200uA,2mA,20mA | 200uA,2mA,20mA,100mA | ||||||
Distortion | Sine waveform | |||||||
Distortion | ≤2%(No load or pure resistance load) | |||||||
Crest factor | 1.3~1.5 | |||||||
Waveform mode | DDS+amplifier | |||||||
Rise time | 0.3s~999.9s 0=OFF | |||||||
Test time | 0.3s~999.9s 0=continuous test | |||||||
Descend time | 0.3s~999.9s 0= OFF | |||||||
Interval time | 0.0s~999.9s 0= OFF | |||||||
Output mode | N mode,G mode | |||||||
DCW | Output voltage | range | 0.050kV~6.000kV | |||||
accuracy | ±(1.5%read value +5V) | |||||||
resolution | 1V | |||||||
Max output power | 60W(6.000kV/10mA) | 300W(6.000kV/50mA) | ||||||
Max test current | 10mA | 50mA | ||||||
Leakage current range | 2uA,20uA,200uA,2mA,10mA | 2uA,20uA,200uA,2mA,20mA,50mA | ||||||
Ripple factor | ≤5% | |||||||
Discharge time | ≤200ms | |||||||
Maximum charge current | 10mA | 50mA | ||||||
Voltage rise time | 0.3s~999.9s 0= OFF | |||||||
Test time | 0.3s~999.9s 0=continuous test | |||||||
Fall time | 0.3s~999.9s 0= OFF | |||||||
Interval time | 0.3s~999.9s 0= OFF | |||||||
Delay alarm time | 0.3s~999.9s 0= OFF | |||||||
Output mode | N mode,G mode | |||||||
IR | Output voltage | range | 0.100kV~1.000kV | |||||
accuracy | ±(1.5%read value +5V) | |||||||
resolution | 1V | |||||||
Maximum upper-limit | 50.00GΩ | |||||||
Maximum lower-limit | 50.00GΩ | |||||||
Minimum lower-limit | 1MΩ | |||||||
Voltage rise time | 0.3s~999.9s 0=OFF | |||||||
Test time | 0.3s~999.9s 0=continuous test | |||||||
Interval time | 0.0s~999.9s 0= OFF | |||||||
Auto range | ON/OFF | |||||||
Discharge time | ≤200ms | |||||||
Volt meter | Range | AC | 0.050kV~5.000kV | |||||
DC | 0.050kV~6.000kV | |||||||
accuracy | ±(1.5%read value +5V) | |||||||
resolution | 1V | |||||||
Displayed value | RMS | |||||||
Ampere meter | Range | AC | 0 ~ 20mA | 0 ~ 100mA | ||||
DC | 0 ~ 10mA | 0 ~ 50mA | ||||||
resolution | AC | 200uA:0.1uA,2mA:1uA,20mA:10uA | ||||||
DC | 2uA:0.001uA, 20uA:0.01uA, 200uA:0.1uA, 2mA:1uA, 10mA:10uA | 2uA:0.001uA, 20uA:0.01uA, 200uA:0.1uA, 2mA:1uA, 20mA,50mA:10uA | ||||||
Accuracy | ≥2Ma:±(1.5%+5counts),<2mA:±(2%+5counts) | |||||||
OFF SET | The current of the test leads and the accessories can be deducted. | |||||||
Test mode | GND:RETURN terminal connect to the outer case FLOAT:RETURN terminal do not be connected to the outer case | |||||||
OHM METER | Range | 1MΩ~50000MΩ | ||||||
Resolution | 1MΩ~9.999MΩ:0.001MΩ,10MΩ~99.99MΩ:0.01MΩ, 100MΩ~999.9MΩ:0.1MΩ,1000MΩ~9999MΩ:1MΩ 10GΩ~50GΩ:10MΩ | |||||||
Accuracy | 0.100kV~0.200kV:1MΩ~999MΩ ±5%,1000MΩ~3000MΩ ±10% 0.201kV~0.499kV:1MΩ~999MΩ ±5%,1000MΩ~5000MΩ ±10% 0.500kV~1.000kV:1MΩ~999MΩ ±5%,1000MΩ~9999MΩ ±10% 10000MΩ~50000MΩ ±15% | |||||||
TIMER | Range | 0~999.9s | ||||||
Resolution | 0.1s | |||||||
Accuracy | ±(0.1%+50ms) |