1400C heating element for lab furnace SiC heating element

1400C heating element for lab furnace SiC heating element

USD $20 - $150 /Piece

Min.Order:2 Pieces

Supply Ability:
200 Piece / Pieces per Month
Port:
zhengzhou qingdao beijing

Quick Details View All >

Brand Name:
CY
Model Number:
CY-H1400
Place of Origin:
Henan, China

Zhengzhou CY Scientific Instrument Co., Ltd.

Business Type: Manufacturer,Trading Company
Zhengzhou Henan China
Main Products: Lab muffle furnace ,Tube furnace ,CVD system ,Plasma Cleaner ,Dental zirconia furnace

Product Details

1400C heating element for lab furnace SiC heating element

 

Specifications of SiC heating element

1.SiC heating elment 
2.shape:ED,DB,SC,SCR,U,W 
3.diameter:12~55mm 
ing temp:600~1600 celsius 
5.high quality ,long servers

 

1.The General Description of SiC heating element

1). production process:made into blank,silicided under high temperature and re-crystallized.

2). character: high-applied temperature ,anti-oxidization ,anti-corrosion ,long service life ,little deformation,easy installation and maintenance.

3). usage: used in various high temperature electric furnaces and other electric heating devices,such as in the industries of magnet,ceramics,metallurgy ,glass,and machinery.

4). physical properties of element 

specific gravity

2.6~2.8g/cm³

bend strength

>300kg

hardness

>9 MOH′S

Tensile strength

>150kg/cm³

porosity  rate

<30%

 radiancy

0.85

 

2. Advices of using SiC heating element

1). In order to keep furnace temperature and the charges uniform of each element, before installation resistance distribution must be carried.

2). As the element is very brittle ,be careful while install and maintain to avoid any damage.

3). when start to operate the furnace ,raise voltage slowly and gradually,never give full load at one time.

 

Inquiry data should be clear:

a). Outer diameter (OD)

b). Length of Hot zone (HZ)

c). Length of Overall (OL)

d). Resistance (at 1050°c +/-50°c)

 

Dia(OD)

(mm)

Length of hot zone(HZ)(mm)

Length of cold zone(CZ)(mm)

Overall Length(OL)(mm)

Range of Resistance(Ω) (ohms)

8

100-300

60-200

240-700

2.1-8.6

12

100-400

100-350

300-1100

0.8-5.8

14

100-500

150-350

400-1200

0.7-5.6

16

200-600

200-350

600-1300

0.7-4.4

18

200-800

200-400

600-1600

0.7-5.8

20

200-800

250-600

700-2000

0.6-6.0

25

200-1200

250-700

700-2600

0.4-5.0

30

300-2000

250-800

800-3600

0.4-4.0

35

400-2000

250-800

900-3600

0.5-3.6

40

500-2700

250-800

1000-4300

0.5-3.4

45

500-3000

250-750

1000-4500

0.3-3.0

50

600-2500

300-750

1200-4000

0.3-2.5

54

600-2500

300-750

1200-4000

0.3-3.0

 

Installation
When installing heating element into furnaces, a separator should be placed between the two shanks of the heating element before hanging the element inside the furnace chamber.  The proper separator width should be selected to fit the shanks smoothly and avoid damage to the element.  

 

Warning:

Exercise extreme caution to not apply any stress on the elements during installation. Heating elements are very brittle and could break if the separator is too wide or if too much pressure is applied on them. 
No warranty will be issued for damage caused by improper installation.

Inspect your delivery:
Please check package for damages before accepting delivery. 
Should damages found, please keep all packaging materials, take pictures, and report immediately to carrier

No Warranty:
Kindly note this item is a consumable part, and not covered by warranty. Make sure to inspect for defect(s) thoroughly before use. 

Contact Supplier

Ms. Jibin Wang trade manager Chat Now
Telephone
86-371-55199322
Mobile
8613837189935
Fax
86-371-86036875
Address
cuizhu road, hi-tech zone Zhengzhou,Henan

You May Like

New Products

Recommended Products

View More View Less