DDSJ Benchtop Conductivity Meter with Temp, Conductivity, TDS, Salinity
USD $150 - $600 /Set
Min.Order:1 Set
Hangzhou Chincan Trading Co., Ltd.
DDSJ Benchtop Conductivity Meter with Temp, Conductivity, TDS, Salinity
DDSJ-308A
DDS-307A
Feature
1. Easy operation, Large and light screen
2. Lower price
3. DDS-307A : ATC: 0.0~40.0℃(Manual/Auto)
Wide measurement function: Conductivity, TDS and temperature
4. More choosing: Temperature compensation: Manual or Auto
Power: 110V—USA standard or Brazil standard
220V—EU Standard or Brazil standard
5. Measurement error avoided: Electrodes produced by us
Specifications
Model | DDSJ-308A | DDS-307A | DDS-307 | |
Specifications | ||||
Parameters | Conductivity/TDS/Salinity/℃ | Conductivity/TDS/℃ | Conductivity | |
Range | Conductivity | 0.000μS/cm~199.9mS/cm | 0.00μS/cm~100mS/cm | |
TDS | 0.000 mg/L~99.9g/L | 0 ~1000mg/L | — | |
Salinity | 0.00~8.00% | — | ||
Temperature | -5.0~105.0℃ | 0.0~60.0℃ | — | |
Resolution | Conductivity | 0.001μS/cm | 0.01μS/cm | |
TDS | 0.001 mg/L | 1 mg/L | — | |
Salinity | 0.01% | — | ||
Temperature | 0.1℃ | — | ||
Accuracy | Conductivity | ±0.5%(FS) | ||
TDS | ±0.5%(FS) | ±1.0%(FS) | — | |
Salinity | ±0.1% | — | ||
Temperature | ±0.3℃±1 bit | — | ||
Stability | (±0.3%(FS))/3h | (±0.33%(FS))/3h | ||
Temp compensation | Manual/Auto 0.0~50.0℃ | Manual/Auto 0.0~40.0℃ | Manual 15.0~35.0℃ | |
Temp coefficient | Adjustable | 2% | ||
Calibration | May calibrate electrode constant or TDS factor | — | ||
Data storage | 200 data with time-and-date stamp | — | ||
View, delete and print | Yes | — | ||
Display | Backlit LCD | LCD, dual parameter | LCD, dual parameter | |
Output | RS-232 | — | ||
Software | 1.1DC | — | ||
Power | Universal power adapter (9V DC,300mA, inside – positive, outside – negative) | AC (220±22)V,(50±l )Hz | ||
Case outside serial No. | WXS-A001-1 | WXS-A008-1 | WXS-A008-1 | |
Dimension (mm),Weight(kg) | 290×200×70,1 | 300×220×90,1.3 | 300×220×90,1.3 |
Cell constant and correspondent conductive measuring range (DDSJ-308A and DDBJ-350 conductivity meter):
Cell constant (K) (cm-1) | 0.01 | 0.1 | 1 | 10 |
Measuring range | (0.000~19.99) μS/cm | (2.00~199.9) μS/cm | 20.0μS/cm~19.9 ms/cm | 200μS/cm~199.9 ms/cm |
For solutions |
Cell constant and correspondent conductive measuring range (DDS-307A, DDS-307 and DDB-303A conductivity meter):
Cell constant (K) (cm-1) | 0.01 | 0.1 | 1 | 10 |
Measuring range | 0.0~2.0 | 0.2~20.0 | 2~1×104 | 1×104~1×105 |
For solutions (μS/cm) |
This year Shipping shows: