Clen Clenbuterol Rapid Test Kit
Negotiable /Box
Min.Order:10 Boxes
Introduction
Specifications
Principle: Competitive lateral flow immunochromatographic assay
Intended use: Detection of Clenbuterol, Ractopamine and Salbutamol
Specimen: Tissue, urine, serum, feed
Specificity: > 99%
Reading time: 5 - 10 min
Shelf life: 18 months
Storage: 2- 30 °C. No freezing.
Packing size: 10 tests / box
Cat No. | Specimen | Detection limit |
W81100-1 | Tissue | 3, 3, 5 ppb |
W81100-2 | Urine | 2, 3, 5 ppb |
W81100-3 | Serum | 2, 3, 5 ppb |
W81100-4 | Feed | 10, 10, 10 ppb |
Advantages
l Quicking Clenbuterol Test enables the detection with simple sample process.
l Quicking Clenbuterol Test is a quick and easy diagnostic tool that can be performed during a consultation in just 10 minutes.
l Quicking Clenbuterol Test can be stored for 18 months at room temperature, between 2 and 30°C.
Instructions
Kit component
- 10×foil pouches each containing one cassette and one desiccant
- 10×pipettes
- 1×assay buffer
- 2×centrifugal tubes (15 mL)
- Product Manual
Test procedure
- Grind a representative sample to the particle size of fine instant coffee (50% passes through a 20 mesh screen). Weigh out a 1.0 g ground portion of the sample into a centrifugal tube and add 3.0 mL of assay buffer.
- Cover the lid and shake sufficiently for 5 min. Do a centrifugation at 7000 rpm for 1 min. Use the clean supernatant in the assay.
- Remove the cassette from foil pouch and place it horizontally on a flat surface.
- Gradually drip 3 drops of sample extraction into the sample well “S”.
- Interpret the result in 5-10 minutes. Result after 10 minutes is considered as invalid.
Interpretation of results
Positive: Only clear C band appears.
Negative: The presence of both C band and T band.
Invalid: No coloured band appears in C zone regardless of whether T band appears.