blood Reagent of Hepatitis B virus HBsAg ELISA test kit
USD $5 - $6.94 /Box
Min.Order:50 Boxes
Certificate |
GMP |
SFDA |
ISO13485 |
ISO9001:2008 |
SPECIMEN COLLECTION AND PREPARATION |
Serum should be prepared from a whole blood specimen obtained by acceptable medical techniques. Either serum or plasma can be used in this test. Remove serum or plasma from the clot or blood cells as soon as possible to avoid hemolysis. Specimen with extensive particulate should be clarified by centrifugation prior to use. Specimen frozen at -20°C or colder may be used. Avoid repeated freeze thaw.
PRECAUTIONS FOR USERS |
STORAGE OF TEST KITS |
Unopened test kits should be stored at 2-8°C. DO NOT FREEZE KIT COMPONNETS. The microtiter plate should be kept in a sealed bag to minimize exposure to damp air. Use up the reagents as soon as possible after the kit is unpacked.
WORKING WASH BUFFER |
Dilute the 20X wash buffer concentrate with deionized or distilled water 1:20. For example, 5 ml of wash buffer concentrate should be diluted to a total volume of 100 mL with deionized or distilled water.
ASSAY PROCEDURE: |
It is strongly advised to analyze each specimen and controls in duplicate. All the reagents should equilibrate to room temperature before use.
INTERPRETATION OF RESULTS |
1. A run is valid if:
1) The full complement of Blanks, Positive and Negative Controls must be included in each assay.
2) Negative Control values must have an absorbance and ≤0.10 after subtracting the Blank.
3) Positive Control value must have absorbance ≥1.00 after subtracting the Blank.
2. Calculation of Cut off Value (COV)
Mean of the Negative Controls (NCx)* 2.1
If the OD value of the negative control is less than 0.05, it should be reported as 0.05. If it is more than 0.05, it should be reported as the actual OD value measured.
Positive OD reading: ≥Cut-off value
Negative OD reading: < Cut-off value
TESTPERFORMANCEANDEXPECTEDRESULTS |
This assay was standardized against Reference Standards provided from the Reference Laboratory for Immunology Product under the Ministry of Health, China.
Clinical Specificity: The clinical specificity of this assay was determinate by a panel of samples obtained from 2500 healthy blood donors and 300 undiagnosed hospitalized patients. The repeatedly reactive samples and samples confirmed positive with the reference test were not included in the calculation of specificity.