Automatic Biochemistry Analyzer AJ-1266 CE Prove
USD $6,000 - $7,000 /Unit
Min.Order:1 Unit
Anjue Medical Equipment Co., Ltd.
Automatic Biochemistry Analyzer AJ-1266
Feature
Type: Fully Automated, Random Access Biochemistry Analyzer
Throughput: 200 test/hour, up to 280 test/hour, with ISE(optional)
Test Mode: End-point, Kinetics, Fixed time, Immunoturbidimetry
Monochromatic, Bi-chromatic & Multi-standard Analysis, Linearity & Nonlinearity Calibration, Single & Dual Reagents can be used
Test Setting: Programmable for single-parameter, multi-parameter, combined parameter, Batch test
Test Range: Clinical Biochemistry, Turbidimetry, ISE, Protein, Drug abuse
Parameters Setting: Completely open for editing test parameters and calculation parameters
Sample System
Sample Tray: 44 sample positions for primary tubes, test tubes and various sample cups. Whole sample try can be removed and replaced. Any position can e aligned as the emergency test position.
Sample Volume: 3ul-30ul, programmed by 0.1ul step
Sample Handling: Dilute and retest automatically when sample’s concentration beyond limitation
Sample probe: liquid level detection and collision protection
Probe cleaning: Automatic washing for both inside and outside, carry over<0.1%
Syringe: High performance syringe, durable and maintenance free
Optional item: Bar code reader
Reagent System
Reagent Tray: 44 position for Single & Dual reagents, reagent positions can be aligned freely
Refrigeration: Refrigerated reagent compartment, 24hours non-stop cooling with Peltier element maintaining temperature 4-15°C
Reagen bottle volume: 35ml and 20ml, Slope bottom design
Reagent volume: 150ul-300ul, programmed by 0.1ul step
Reagent probe: Automatic liquid level detection function and collision protection
Reagent pre-heating function, inventory checking.
Real time monitoring of reagent residual volume and auto alarm
Probe cleaning: Automatic washing for both inside and outside, carry over<0.1%
Reaction System
Reaction tray: 48reaction cuvettes with 6mm optical path, good UV penetrability, low consumable, low cost
Reaction technology: Back-Dividing-Light Technology
Reaction volume:250ul-300ul
Temperature: 37°C,with+0.1°Cfluctuation
Mixing system: Independent mixer
Cleaning: Automatic cuvettes cleaning and complete drainage to reduce carry over
Water consumption: 5L/hour
Liquid detection: Detergent, distilled water, waste liquid automatic detection and alarm
Optical System
Light source: Halogen lamp 6V/10W, life span 3000hours
Wavelength: 340nm,405nm,492nm,505nm,546nm,578nm,620nm,670nm,700nm
Wavelength accuracy: ±2nm
Detector: Photoelectron Diode Array
Absorbance range: 0.000 Abs – 3.000 Abs
Calibration & Quality Control
Calibration mode: Linear (one-point, two-point and multi-point), Logit-Log4P, Logit-Log5P, Spline, Exponential, Polynomial, Parabola
QC rules: Westgard Multi-rule
QC mode: Intra day & Inter day QC
Operating System
Computer: Windows 2000 System or above
Date management: Large capacity for results and reaction curves to be displayed, saved and printed
Report: Edit and print complete reports, format editable, support network printing
Port: RS-232
Working Conditions
Power: 220V±22V, 50/60Hz±1Hz
Power consumption:≤400VA
Environmental temperature: 15°C-30°C
Humidity: ≤85%
Dimension: 760(L)x560(W)x1080(H)mm
Weight: 85Kgs