Product Details

wearable devices ECG tester

Brand Name N/A
Place of Origin China
Model Number WECG400

Product Features

WECG400

The first parallel testing ECG tester specifically designed for wearable devices can simultaneously test ECG function on up to 4 devices with a single computer to accelerate testing time.

It provides compliance tests conforming with IEC 60601-2-47, YY0885, and YY 9706.247-2021, and can output 7 types of standard ECG waveforms to assist wearable devices in meeting regulatory requirements of accuracy.


SPECIFICATION

Standard ECG waveforms

             

Parameters Range Resolutions

ECG Waveform Frequency (Hz) 0.02 ~ 6Hz 0.01

Frequency (BPM) 1 ~ 360BPM 1

Amplitude (-5) ~ 5mV 0.01

Sine Frequency (Hz) 0.05 ~ 500Hz 0.01

Frequency (BPM) 3 ~ 30000BPM 1

Amplitude (-10) ~ 10mV 0.01

Triangle Frequency (Hz) 0.05 ~ 500Hz 0.01

Frequency (BPM) 3 ~ 30000BPM 1

Amplitude (-10) ~ 10mV 0.01

Square Frequency (Hz) 0.05 ~ 500Hz 0.01

Frequency (BPM) 3 ~ 30000BPM 1

Amplitude (-10) ~ 10mV 0.01

Exponential Frequency (Hz) 0.05 ~ 5Hz 0.01

Frequency (BPM) 3 ~ 300BPM 1

Amplitude (-5) ~ 5mV 0.01

Rectangle Pulse Frequency (Hz) 0.05 ~ 5Hz 0.01

Frequency (BPM) 3 ~ 300BPM 1

Amplitude (-5) ~ 5mV 0.01

Pulse width (frequency: 0.05Hz) 1 ~ 300ms 1

Pulse width (frequency: 5Hz) 1 ~ 200ms 1

Triangle Pulse Frequency (Hz) 0.05 ~ 5Hz 0.01

Frequency (BPM) 3 ~ 300BPM 1

Amplitude (-5) ~ 5mV 0.01

Pulse width (frequency: 0.05Hz) 1 ~ 300ms 1

Pulse width (frequency: 5Hz) 1 ~ 200ms 1

 


Input Impedance / Frequency Response Signal add-on

             

Parameters Range Default Values Resolutions

Play raw data Sample rate Maximum: 40K Hz - -

DC offset (non-Variable) -300, 0, +300mV 0mV 300

DC offset (Variable) (-1000) ~ 1000mV 0mV 100

Input impedance switch 100Ω (short circuit) - -

620kΩ / 4.7 nF on/off Off (on: short circuit) -

10MΩ / 180 pF on/off Off (on: short circuit) -

 


Product System Capacity and Standard Requirements Table

Parameters Specifications System Capability

Main output voltage accuracy ±1% for amplitudes of 0.5mVpp or higher ±0.3%

Main output voltage resolution

(DAC resolution) N/A 0.15µV

Frequency / pulse repetition rate accuracy ±1% ±0.1%

Pulse duration / timing accuracy ±1ms ±0.2ms

Resistor tolerance ±1% ±0.5%

Capacitor tolerance ±5% ±5%

Precision 1000 : 1 divider ±0.5% ±0.3%

Sample rate 5kHz ± 0.1% 10kHz ± 0.05% (50ppm)

DC offset

(fixed, noise-free, sourced from the internal supercapacitor) 300mV ± 1% 300mV ± 0.1%

DC offset

(variable, up to 1000mV) Setting ±1% or ±3mV Setting ±1% or ±3mV

Environment 15 ~ 30°C (by design)

30 ~ 80% RH (by design) 5 ~ 40°C

30 ~ 80% RH

altitude<2000M

Overvoltage CAT II