Semi-Auto Biochemistry Analyzer KD720
- Features
- Analytic Modes: kinetic, fixed time, end-point, absorbance
- Wavelengths: 7 standard filters: 340, 405, 492, 510, 546, 578, 630 nm, 1 more filter optional
- Full open reagent system
- Items can be programmed, number of items not limited
- Voice prompting, making the operation interesting
- Display: LCD with high brightness
- Input: Touch screen operation
- Real time graph display and print
- Lamp saving feature, longer lamp lifespan
- Technical Specification
Assay Method |
Reaction rate, end point, two-point, double wavelengths, coefficient, multiple standards, blank reagent, blank sample, absorbance at end point, QC, QC graph and turbidity |
Temperature System |
PELTIER temperature control component, 25℃, 30℃, 37℃, accuracy: ±1℃ |
Light Source |
6V 10W halogen lamp of more than 2000 hours lifespan |
GML Precision |
±2 nm |
FWHM |
≤10 nm |
Stray Light |
≤0.5(absorbance≥2.5) |
Absorbance Range |
-0.214-3.000 |
Absorbance Linearity |
±1% |
Absorbance Repeatability |
±0.001 |
Precision |
≤1% |
Carry Over |
≤1% |
Resolution |
1mAbs/h |
Stability |
≤3mAbs/h |
Flow Cell |
Permanent quartz flow cell, minimal volume is 32uL, and disposable flow cell is selectable |
Sampling System |
Pressure auto releasing pump ensure the accuracy of suction volume |
Absorption Volume |
200uL-3000uL |
Sample Volume |
10uL-50uL |
Display |
240*128 LCD, touch screen |
Interface |
RS232 Serial port |
Working condition |
Temperature 10-30℃, RH: 20-80% |
Voltage |
AC85-264V, auto adapt |
Power |
≤100W |
Quality System |
CE |
- Product Show