WET-2000E Particle Shape and Particle Size Analyser
USD $1 - $10000 /Set
Min.Order:1 Set
Gold APP Instruments Corporation Limited
WET-2000E laser particle size analyzer adopts wet dispersion principle, it is most economical and popular for production QC. Laser diffraction particle size analyzers are used to measure the sizes of particles in a material. Particle size is calculated by measuring the angle of light scattered by the particles as they pass through a laser beam. Laser diffraction analyzers are used in many applications, including manufacturing, quality control and product development.
WET-2000E Particle Shape and Particle Size Analyser Specifications
Model Name | WET-2000E | |
Standards | ISO13320-1:1999, GB/T19077.1-2008, Q/JWN001-2009 | |
Measuring Range | 0.1μm-300μm | |
Channels Number | 39 | |
Accuracy | <1% (national standard sample D50) | |
Repeatability | <1% (national standard sample D50) | |
Light Source | High performance semiconductor laser (λ= 632.8nm, P>2MW, life >25000hour) | |
Dispersion Method | Ultrasonic | Frequency:40KHz, power:35W, time: ≥1S |
Stir | Revolutions speed: 0-300RPM (adjustable) | |
Circulate | Flow rate: 8L/min, Power Rate:10W | |
Sample Pool | Volume: 350mL | |
Micro-sample Pool | Volume: 10mL (optional) | |
Operation Modes | Embedded display + external computer | |
Optical Calibration System | Manual | |
Test Speed | <2mins for each time | |
Dimension | L66cm×W32cm×H40cm, 25Kgs |
WET-2000E Particle Shape and Particle Size Analyser Features:
1. By Mie scattering theory
2. Fully built-in integrated dispersion system contains ultrasonic stirring, ultrasonic dispersion and cycling system.
3. Unique unconstrained free fitting technology, make particle analysis not restricted by any functions, truly reflect particles distribution, ensure the good accuracy.
4. Converging light Fourier transform light path, efficiently improve resolution ratio of sub-micron particles.
5. Main & auxiliary detectors are log-shaped, guarantee the effective capture of the scattered signal.
WET-2000E Wet Laser Particle Size Analyzer Schematic
WET-2000E Particle Shape and Particle Size Analyser Functions:
1. Analysis Modes
Free distribution, R-R distribution, logarithm normal distribution, mesh number classification etc.
2. Statistic Method
Volume distribution and quantity distribution
3. Statistic Comparison
Several testing results of samples;
Different batches of samples testing result;
Samples before and after processing;
Test result of samples from different time.
4. User-defined Analysis
Figure out percentage according to the particle size;
Figure out particle size according to the percentage;
Figure out percentage according to the particle size range;
Meet demands of representation of particle test in different industries.
5. Test Report
Word, Excel, Photo (bmp), Text etc.
6. Multi-language Support
Chinese & English (others are available)
7. Intelligent Operation Mode
Automatically control water inflow, dispersion, and analysis. Easy-to-use software and fully automated operation to minimize training requirements and release operator time.
Operation Interface:
1. Operation interface:
Run the software, access control systems interface, manually start the sample testing.
2. Testing analysis interface and report
After testing be completed, according to requirement to select records, the average result will be calculated, system generate analysis records form. When test in automatic mode, without data processing, the system automatically gets analysis report and save the test records after a comprehensive analysis.